ALTOSONIC V
3
Ultrasonic Measurement Principle
Design specifications
Performance
Calibration
Installation
Diagnostics
Ultrasonic Measurement Principle
Introduction to
Measurement Principle
-
Liquids
5
Ultrasonic Measurement Principle
Design specifications
Performance
Calibration
Installation
Diagnostics
UFS-V Flow Sensor Design Specifications
Diameters:
6”, 1500 lbs 4” - 40” (100 -1.000 mm)
Material :
Stainless steel 316/316L
12”, 150 lbs
Duplex, SMO and others
on request
Pressure classes:
150 lbs, 300 lbs, 600 lbs
20”, 150 lbs
up to 1.500 lbs on
request
7
UFS-V Flow Sensor Design Specifications
8
Ultrasonic Measurement Principle
Design specifications
Performance
Calibration
Installation
Diagnostics
Altosonic-V Performance
10
Altosonic-V Performance
Correction for flow profile disturbances
11
Altosonic-V Performance
5 Beams ...
Provide more information on
the flow profile
Compensate for changes in
flow profile
Better performance with non
Position of 5 beams
symmetrical flow profiles
Unique configuration of Good reproducibility
5 parallel beams
High accuracy ±0,15%
Based on stable calibrated over wide Reynolds range
geometry
Create redundancy
Extensively tested
Symmetrical
used for flow profile check
13
Altosonic-V Performance
14
Altosonic-V Hardware Modules
RS 485 RS 485/422
UFS-V UFC-V UFP-V
for hazardous
areas
UFP-V 19” rack
version for non
hazardous areas
15
UFS-V Flow Sensor Design Specifications
Diameters:
6”, 1500 lbs 4” - 40” (100 -1.000 mm)
Material :
Stainless steel 316/316L
12”, 150 lbs
Duplex, SMO and others
on request
Pressure classes:
150 lbs, 300 lbs, 600 lbs
20”, 150 lbs
up to 1.500 lbs on
request
16
UFC-V Signal converter
Converters
• Control 5 measurement paths
• Perform actual flow measurement of
5 channels
• Raw data of each individual path
sent over via RS 485
• Processed by UFP-V flow
processor
17
Ultrasonic Flow Processor (UFP)
UFP-V Field version
For hazardous areas
Identical to UFC-V cover cabinet
Copper free aluminium
Communication
Output via RS 485 / 422
18
Altosonic-V Measurement Functionality
Standard:
Actual volumetric flow rate
Actual volumetric flow total
Status & diagnostic information
Velocity of sound
Flow direction (forward/reverse)
Body temperature measurement
Optional :
Actual volumetric flow rate
Line pressure and temperature input
Densitometer input (frequency / 4-20 mA )
Viscosity transmitter input (4-20 mA)
19
Altosonic-V Measurement Functionality
API 2540
ASTM-IP (D1250, 1953 ed.)
LPG (GPA) TP 25
ULHC (Special application for Russia)
20
Ultrasonic Measurement Principle
Design specifications
Performance
Calibration
Installation
Diagnostics
Altosonic-V Certification & Approval
Certificate of compliance
with OIML R 117, class 0.3
“Measuring systems for liquids other than
water” (edition 1995)
Issued by NMi
Complete test program & report available
Basis for national type approval
certifications by national board of Weight
& Measures
22
Altosonic-V Certification & Approval
23
Altosonic-V Certification & approval
API Standard on Ultrasonic Flowmeters
Requirements Performance
API standard on UFM ALTOSONIC V
Accuracy: not Easily within Uncertainty
specified
Repeatability: not ± 0,027% band
specified conform API
Uncertainty: specifications
± 0,027% uncertainty
(with 95% confidence level)
For example 5 runs:
repeatability 0,05%
No certificates of compliance
with API issued
25
Altosonic-V Certifications & approval
CT Requirements for Offshore
Design specifications
Performance
Calibration
Installation
Diagnostics
Altosonic-V Calibration
Standard calibration
At 5 flow points
(10, 20, 50, 70, 90%)
5 times repeat
With viscosity
comparable to
application:
min 1 cSt, max 120 cSt
KROHNE calibration
certificate standard
provided
28
Altosonic-V Calibration
Optional Calibration to OIML R 117 & NPD
OIML R 117
Calibration at 6 flow points
over operating range
Each point repeated 3 times
29
Ultrasonic Measurement Principle
Design specifications
Performance
Calibration
Installation
Diagnostics
Altosonic-V MID certified data
Sizes 4 to 24 inch (Bigger sizes
available)
Process temperature -200 to +250 ºC
Viscosity range 0.1 to 400 cSt (For all diameters)
Flow range 0.2 to 10 m/s (turndown 1 : 50)
Ambient temperature
UFS-V -25 to +55 ºC
UFC-V -40 to +70 ºC
31
Altosonic-V MID certified data
Explosion safety ATEX, Zone 1
UFS-V Eex ib IIC T6
UFC-V EEx d [ib] IIB T5
UFP-V EEx d [ib] IIB T5
Protection class
UFS-V IP 67
UFC-V IP 55 (/IP 65/67)
32
Altosonic-V Installation
Hazardous Area
UFS-V II 2 G EEx ib IIC T6 ..T4
34
Ultrasonic Measurement Principle
Design specifications
Performance
Calibration
Installation
Diagnostics
Altosonic-V Long term stability
Verification of a 20” Altosonic V over period of 10 years
Test data:
Instrument 20” Altosonic V
Test Facility Verifications at SPSE
Test medium Oural (5 cSt)
Time interval Period of 10 years (1999-2009)
Test results:
Difference 1999 – 2009: 0.08%
Results include installation effects
No maintenance has been done to the meter
Conclusion:
Very stable performance of 20” Altosonic-V.
No systematic shift over period of 10 years. Average variation lies
Between 0.08%. Variation is caused by the reproducibility of the flow
meter,calibration and installation effects.
36
Altosonic-V: Long term stability
37
Altosonic-V Long term stability
ALTOSONIC V for Offshore Flow Metering
Installation Statoil Vigdis/Snorre Crossover (1997)
• Customer:
Saga (now Statoil)
• Application: Oil transfer between
Snorre and Vigdis process trains
• 2 x 8” ALTOSONIC V (300 lbs)
in master / duty set-up
• Periodically verified against each
other (e.g. scaling effects)
• No permanent prover
installlation, No filters
No maintenance done • NPD acceptance for the metering
since their installation in 1997 solution
38
38
Altosonic-V Long term stability
39
Altosonic-V: Long term stability
40
Altosonic-V Long term stability
41
Altosonic-V Long term stability
Field Trial at Dalmeny Tank Terminal BP, UK
Trial Project
Trial period of 15 months
Strainer removed from meter run
Line was left unheated, without insulation
Goals:
Establish if meter / transducers suffer from
wax or deposition
Verify accuracy, repeatability & long term
stability
Same proving criteria as for turbines:
> 5 consecutive runs within ± 0.05%
repeatability,
> within 10 runs by bi-directional prover
> or else fiscal status of flowmeter not valid
42
Altosonic-V Long term stability
Long term stability, No maintenance
Results: BP Dalmeny
No significant long-term drift or
scaling observed after 15
months
Flowmeter reliable & repeatable
all prove results within +/-
0.10% of original k-factor
Meter body showed no sign of
wax or deposition.
No intervention for maintenance
or calibration during 15-month
Internal ALTOSONIC V flow sensor trial period
after 1 year operation. Sludge present
after 1 year did not influence
measurement nor its performance
43
Altosonic-V Long term stability
Long term stability, No maintenance
Conclusion BP
• Trial was a success
• No impact on nor disruption
to, routine operations
• No adverse impact on FPS
customers
44
Altosonic-V: Long Term Reliability
Maintenance-free
46
Ultrasonic Measurement Principle
Design specifications
Performance
Calibration
Installation
Diagnostics
Altosonic-V Diagnostics
ALTOSONIC V: Integrated Diagnostics
48
Altosonic-V Diagnostics
ALTOSONIC V Integrated Diagnostics
49
Altosonic-V Diagnostics
Integrated Diagnostics
Several self tests performed Diagnostics of direct environment
Memory checks Flow velocity per path
Parameter validity check Sound velocity per path
Plausibility checks on
individual measurements Provide indication of
Validity check of measured Scaling
values Presence of higher viscosities,
Sound velocity check i.e. entrained air, gas or wax
comparison between 5 Warning that measurements may
beams not be within specified accuracy
class (measurement continues!)
50
50
Altosonic-V Diagnostics
Check on Scaling
Path 1
Path 1 2 3
51
Altosonic-V Diagnostics
Indication of Indication of
Too high contents of solids or Obstruction in the flow
particles Damaged valve
Too high gas contents e.g. Observation
resulting cavitations Velocity profile disturbances
Observation Diagnostic feature: Visualization
Transmitted signal from one of the Flow Profile
sensor not correctly received by
the other sensor.
Alarm: Path failure
Automatically an alarm is
generated
Real time correction of missing
values based on historical data
Altosonic-V Summary
• 5 measurement channels
• For maximum performance
• Accuracy: ± 0.15%of MV
• Repeatability: ± 0.02%of MV
• Compliant with
OIML R117 & NPD & API
• Measurement independent of
viscosity
• Only true multi product
flowmeter on market
• No k-factor setting after product
change
53
Altosonic-V Summary
54
Altosonic-V
55
Thank You