Professional Documents
Culture Documents
AKTA Purifier Core 100 Training
AKTA Purifier Core 100 Training
AKTA Purifier Core 100 Training
System Features
Flowrate: 0.01 100 ml/min System Pressure: 0 10 MPa Conductivity Range: 1uS/cm 999.9 mS/cm pH Range: 0 14 (+/-0.1 pH unit) UV Measurement: 190 700 nm in 1nm steps (3 Simultaneous)
UV
Injection valve (V1) 1 2 3 7 4
3 Wavelength detector
Cond
6 5
Conductivity monitor
pH
pH detector
Sample
Application of Sample
Injection Valve Loop Loading and Rules Superloop Direct Injection
Collecting Peaks
Flowthrough Fractionation Elution Fractionation Peak Fractionation Delay Volume
Buffer OUT
Ball outlet check valve
Check for leakage. If there is a sign of liquid leaking between the pump head and the housing side panel or increased or decreased volume of rinsing solution, replace the piston Seals.
When changing eluent, it is important to remove trapped air. Backwash pump diaphragm Ball inlet check valve
Wash IN
Buffer IN
The purge line is recycled back in the resevoir. Make sure that the resevoir lines are secured.
Syringe fitting
Wash IN
Buffer IN
Proportion Valve A2 A1 B2 B1 A B A B
The Mixer volume is best changed for different flow rates. When a command for a %B mix is received by the Pump it takes 3-4 X the mixer volume before it is achieved COLUMN TOP 0%
Mixer, 2mls
COLUMN TOP
KTA 10 mls/min KTA 100 mls/min GRADIENT <6 mls/min 6 -14 mls/min <30 mls/min 30 - 50 mls/min 50 - 100 mls/min MIXER 0.6 ml mixer 2 ml mixer 2 mls mixer 5 mls mixer 12 mls mixer
20%
This appears in the program as System compensation volume and Gradient delay volume
Seal
Buffer OUT
Only manually purge when pump contains large volume of air Unscrew valve and the seal moves out of the way. Buffer can then be sucked through the Pump. Note the check valves opening.
Air
Buffer IN
Pump A
Pump B
Common inlet
A2
A1
Common inlet
B2
B1
Manual A1 or A2
Flowpath BufferValveA
Manual B1 or B2
Flowpath BufferValveB
Execute
Open
ON-LINE FILTERS
Polymer Fibre Filter ( 5um pores ) Support Mesh ~ Stops fibres from entering the chromatography steam.
P-900 Pump High Flow Fibre Filter ( for Soft Gel Columns ) 18-1027-11 ( 10/pack)
10 0u
Injection valve V1
Accumulator (40ul)
1. Suppresses Air Bubbles 2. Stops Siphoning Effect on Buffer 3. Closes Pump Check Valves efficiently
Conductivity Flow cell (14ul) FR-902 FR-904 (0.2Mpa, 5ul) (0.4MPa, 7ul)
Column Pressure
TOP ~ Full resistance reflects the addition of the Flow Restrictor. BOTTOM ~ Resistance to flow, due to Flow Restrictor.
2 90 RF 0.5
0.3
MPa
TOP ~ Full resistance to flow, by gel bed. BOTTOM ~ No resistance to flow, open to atmosphere. This means you should ADD the 0.2MPa to the recommended Column Pressure, ie. HiTrap (0.3 +0.2)= 0.5MPa Pressure gradient down column (0.5 0.3) Pressure difference = (0.3 0.0) = 0.3 MPa
Column Pressure
WARNING ! However the set pressure must not EXCEED the recommended Column Hardware pressure.
Column Pressures
Storage Reservoir
1:1 mixture Buffer pH 4 and 1M KNO3
SAMPLE APPLICATION
Different ways of applying sample
Waste 1 2 4
Waste
LOAD
COLUMN 7 3 5 6 PUMP
Waste
Waste
COLUMN
INJEC T1
2 3 4 5
7 6
PUMP
Waste
Waste
WASTE
End
Beginning
200ul Load
Partial Loading Loop = No more than 1/2 Loop Volume Total Loading Loop = Inject 2X Loop Volume
System Pump
LOAD
Syringe Needle Adaptor Sample Loop
Waste
While buffer is pumped through INJECTION VALVE to column Sample can be pushed ( LOADed ) into Loop
7 3 4 5 6
System Pump
Sample Loop
Waste
INJECTION VALVE changes from LOAD to INJECT Buffer pushes ( INJECTS ) sample onto column
INJECT LOAD
System Pump
6
SuperLoop
10 0
Waste
INJECTION VALVE Changes from LOAD to INJECT to column While buffer is pumped through INJECTION VALVE System PumpsSuperloop Syringe Loads (injects) Sample from Superloop to column
LOADING SAMPLE LOOP with SAMPLE PUMP ( sample volumes >10 mls )
Injection Valve INV-907 IV-907
1 2 7 3 4 6
5
System Pump
LOAD
Sample Loop
Sample
NOTE: Sample Does not enter pump This means Minimal wastage
Sample Pump
Waste
LOADING SAMPLE LOOP with SAMPLE PUMP ( sample volumes >10 mls )
Injection Valve INV-907 IV-907
1 2 7 3 4 6 5
System Pump
Sample Loop
Sample
NOTE: Sample Does not enter pump This means Minimal wastage
Sample Pump
Waste
INJECT LOAD
System Pump
6
SuperLoop
Sample
NOTE: Sample goes throughBuffer pushes NOTE: the pump and is onto the Samplepushed into superloop column
Waste
Sample Pump
DIRECT LOADING SAMPLE with SAMPLE PUMP ( sample volumes > 10 mls )
Injection Valve INV-907 IV-907
1 2 7 3 4 5 6
LOAD INJECT
System Pump
Extra Tubing
Sample
Waste
Sample Pump
LOAD INJECT
7
Air Sensor
(watch)
System Pump
6
3 4 5
Sample
DIRECT LOADING SAMPLE with SYSTEM PUMP ( sample volumes > 10 mls )
1 2 7 3 4 5 6
LOAD
System Pump
Inlet A2
Buffer
Sample
LOAD
System Pump
8 7 6
Inlet A1
Change To Buffer
A12
2 3 4 A13 5
Sample
Collecting Peaks
Different ways of collecting Fractions
0%B
100%B
%B 100
Tube
UNBOUND
Inject
UV Watch
250
Slope or Level
Start
50
End
X1
UV280 Conc %B
X2
X3
A1 A2 A3
Flowthrough Fractionation
Eluate Fractionation
Fixed volume
Peak Fractionation
Waste
FRACTION DELAY
Delay volume
START FRAC END FRAC
Waste
A1
A2
A3
A4
A5
Waste
Fraction Collector
vo lu m e De la y
A1 A2 A3 A4
Delay volume continues collecting fractions after end of run until the delay volume is achieved
Optional Functionality