Professional Documents
Culture Documents
Lecture 20-03-12
Lecture 20-03-12
CLINICAL CHEMISTRY
INSTRUMENTS &
PRINCIPLES
BY AZEEMA JAMIL
INSTRUMENTS CAN BE DIVIDED ON THE BASIS OF THEIR
PRINCILPES.
A- SPECTROPHOTOMRTY
Following instrument based on spectrophotometer principle.
1- Genesis 2- Dimension 3- ELISA
4- Hitachi 5- Synchron 6- Micro lab
7- Advia 1800
B- ISE METHOD
Following instrument based on ISE principle.
1- Medica Easylyte
2- Synchron (CX3-portion)
C- CHEMILUMINESCENT
Following instrument based on CHEMILUMINESCENT principle.
1- ELECSYS 2010/ E-170
2- IMMULITE
D- DIFFERENT TECHNIQUES
Hopefully, you will use the same size curvets for all your measurements,
so b remains the same for all measurements.
Also, since molar absorptivity does not change either, it remains the same
for all measurements.
Because Beer’s Law states that absorbance (A) is directly proportional to concentration. the
following mathematical relationships can be established
Aunk
Cunk x Cstd
Astd 5
The following simulation illustrates the procedures for making spectrophotometric
measurements.
• First, the intensity of light (I0) passing through a blank is measured. The intensity is
the number of photons per second.
• The blank is a solution that is identical to the sample solution except that the
blank does not contain the solute that absorbs light. This measurement is
necessary, because the cell itself scatters some of the light.
• Second, the intensity of light (I) passing through the sample solution is measured.
• Third, the experimental data is used to calculate two quantities:
• the transmittance (T) and the absorbance (A).
T= I/IO
A = - log10 T
Transmittance; The fraction of light that passes through the sample and reaches the
detector. The remainder of the light is the fraction of the light absorbed by the
sample.
• Absorbance; A logarithmic measure of the amount of light absorbed (at
particular wavelength) as the light passes through a sample or substance.
There are two types of photometric assay:
1- Oxidation method
2-Reduction method
Reduction is the gain of an electron. Sometimes we also have H
ions along for the ride, so reduction also becomes the gain of H.
Synchron
Cx3-portion (Na,K,Cl)
MEDICA EASYLYTE
ION-SELECTIVE ELECTRODE (ISE)
This principle is used for the determination of sodium ion, potassium
ion, and chloride ion concentration by measuring electrolyte activity in
solution.
Sample is mixed with a buffered solution.
The buffered established a constant ionic strength and set a constant
activity co efficient for the electrode.
Sodium -> consist of glass electrode
Potassium -> consist of electrode with valinomycin membrane
Chloride -> Ag/ Agcl type.
Lithium -> Plastic tube electrode
When sample / buffer mixture contact with their specific electrode, the
ion exchange take place and a change in potential (voltage), is
developed at the face of the electrode and change in potential allows the
calculation of each analyte in a solution.
CHEMILUMINESCENT
PRINCIPLE
CHEMILUMINESCENCE
SENSITIVITY
Lower detection limits when compared to absorption or
fluorescence techniques in pursuit of the same analyte.
Selectivity
The analyte of interest generates its signal often in the presence of
normally interfering compounds and do not themselves produce light
until the chemiluminescent reagents are mixed together.
LOPHINE
LUMINOL
OXALATE ESTERS
RUTHINIUM TRIS BIPYRIDINE
LUCIFERIN
H2O2 or organic peroxides are commonly used as the oxidants in
the solution phase in chemiluminescent techniques. Hydrogen
peroxide (H2O2) is the simplest peroxide and an oxidizer.
The bond between the two oxygen atoms is easily cleaved, and
releases a significant quantity of energy when broken.
ELECTROCHEMILUMINESCENCE
23
FREEZING POINT
Temperature at which a liquid solidifies under a specified pressure .
The temperature at which the liquid and solid phases of a substance of
specified composition are in equilibrium at atmospheric pressure.
Fast Cool:- The thermistor probe (sensing the temperature of the sample) and the stir
–wire (vibrate the sample) dip in the sample, the sample begin cooling rapidly.
Slow Cool:- The temperature of the sample reaches O0C, the cooling rate of the
sample slow.
Freeze:- The sample has been sufficiently super cooled and sample become freeze.
Heat of fusion:-The heat of fusion escape the sample and the temperature rises to
the actual freezing point of the sample
Read out:- As the plateau is constant it locks the reading onto the display.
ELECTROPHORESIS
The rate at which different molecules move along the media will
vary depending on the physical characteristics of the molecules.
28
ATOMIC ABSORPTION
30
Competitive Immunoassays (RIA,MEIA,FPIA)
Competition between tagged and un-tagged antigen for limited
antibody
Tagged antigen Reagent
Untagged antigen Patient antigen we want to measure
Specific antibody Reagent
Let the competition begin !!!
Mix the three components together
Allow the antigens to compete for the limited antibody
Antibody will bind with tagged or un-tagged antigen ( it
doesn’t care )
Separation Step : Antibody-Antigen complexes are
separated from free antigen
Tagged antibody-antigen complex is measured
31
The tagged antigen and antibody from the reagent kit are
constant.
The only variable is the concentration of the patient antigen
( the thing we want to measure )
A standard curve can be constructed with known antigen
concentrations giving the following general results
Analyte-antibody complexing
2
Adding sample increases scatter slightly
1
Fluid in the cell has causes little scatter
Time
1. Buffer addition 2. Sample Addition 3. Antibody addition
Nephelometry can be used to detect either antigen or antibody.