Ion chromatography 1
CHROMATOGRAPHY
Ion Chromatography
is a chromatography process that separates
ions and polar molecules based on their
affinity
Ion
Chromatograph
y
4
Ion chromatography
HPLC (UHPLC) IC
INTRODUCTION Sample 5 – 20 µL Variable volume until 1000 µL
System Stainless steel, 0 – 50 (100) MPa, typical 15 – PEEK, 0 – 35 MPa,
25 (90) MPa typical 10 – 20 MPa
Eluent Organic solvents, Ultrapure water with salt,
typical gradient profile typical isocratic method
Separatio Adsorptive effects Chemical reactions
n fast with very sharp peaks typical 15 min for 7 peaks
Detection UV Conductivity
• HPLC VS IC Specialty - Suppression
•
Setup
of IC
Ion chromatography
General setup of an IC
Detector
Mobile phase
(Eluent)
Injection
valve
High pressure
pump
Separation
Separation
column
column
7
•
Eluent in
IC Pemisahan terjadi pada fase diam di dalam kolom kromatogram
kolom
Ion chromatography
Cations
tartaric acid/dipicolinic acid
CONTOH ELUEN
tartaric acid
citric acid
sodium dihydrogene phosphate
oxalic acid/ethylene diamine/acetone
borate/Gluconate
• UV/Vis applications
reversed-phase columns according to pharmacopoeia
Ion chromatography
Metrosep C4 – 100/4.0
Temp
25°C
Injection
5 µL
Detection
Conductivity
Ion chromatography 11
Ion
chromatography
Preparation
IC 940
1. Labu Ukur 250 ml, 1000ml, 2000 ml
2. Pipet Volume 10ml, 15ml, 20ml, 25ml, 30ml
3. HNO3 2.1 mmol/L, Dipicolinic 1.2 mmol/L, Acetone 5%
4. NaCl Pa, CaCl2 Pa, KCL Pa, Na Laktat Pa,
5. Sample RL
Ion
chromatography
Preparation
BM HNO3 : 63 BM : 167,122
V : 2000 ml 2L
ρ HNO3 : 1,39
M : 1,2 mmol/L 0,0012 M
% HNO3 : 65
massa(gr): ?
M1.V1=M2.V2
massa : 401,0928 kg 0,401093 gr
M1 65%= 14,34127 Molar
V1= 15 ml
v2= 250 ml
M2 ? 2
M2= 0,860476 M
860,4762 Mmol/L
L
I
Perhitungan M1.V1=M2.V2
T
Eluen
V1= 2000 ml
Aceton 5%
M1= 2,15 Mmol/L
V1.M1=V2.M2
E
M2= 860,4762 Mmol/L
V1= 2000 ml R
V2= ?
M1= 5%
V2= 4,997233 ml
M2= 99,99 %
V2= ?
V2= 100,01 ml
Ion
chromatography
Preparation
IC 940
IC 940
1. Pipet 20 ml sample RL ke dalam 250 ml ad sampai tanda batas
Preparasi
Larutan
Sample
IC 940
1. BASELINE/STAR
HW
2. INJECT
SAMPLE/STD
3. PEMBUATAN KURVA
STD
4. PEMBACAAN
SAMPLE
IC 940
1. BASELINE/STAR
HW
IC 940
(BASE
LINE)
1. BASELINE/
- Pilih Equilbrium,
- Methode: Rl-3DU
- Start HW
-Tunggu Sampai stabil
30 menit
IC 940
(BASE
LINE)
1. BASELINE/
- Pilih Equilbrium,
- Methode: Rl-3DU
- Start HW
-Tunggu Sampai stabil
30 menit
IC 940
(INJECT
SAMPLE)
1. Pilih Single
Determination
- Setting Komponen
-Setting Komponen
- Setting Komponen
1. Pilih DataBase
2. Klik Kanan Pada
STD yg sudah di
proses
3. Klik Reproses
IC 940
(Pembuatan Kurva)
- Setting Komponen
1. Pilih Components,
klik components
table,klik Na
2. Arahkan kursor di
Peak paling kiri
3. Pilih Update
retention time
4. Pilih Update
IC 940
(Pembuatan Kurva)
- Setting Komponen
1. Pilih Components,
klik components
table,klik K
2. Arahkan kursor di
Peak Tengah
3. Pilih Update
retention time
4. Pilih Update
IC 940
(Pembuatan Kurva)
- Setting Komponen
1. Pilih Components,
klik components
table,klik Ca
2. Arahkan kursor di
Peak Paling Kanan
3. Pilih Update
retention time
4. Pilih Update
IC 940
(Pembuatan Kurva)
- Setting Komponen
1. Pilih reprocessing
2. Pilih From Std of
Reprocessing Table
IC 940
(Pembuatan Kurva)
-Setting Event
1. Pilih Integration
2. Pilih Events
3. Klik No.1 arahkan
kursor ke Peak paling
kiri, batas warna
merah, cek RT dan
masukan kedalam
Tabel (start)
IC 940
(Pembuatan Kurva)
-Setting Event
1. Pilih Integration
2. Pilih Events
3. Klik No.2 arahkan
kursor ke Peak ke
batas warna merah,
cek RT dan masukan
kedalam Tabel (Ends)
IC 940
(Pembuatan Kurva)
-Setting Event
-Setting Event
-Setting Event
1. Cek Kurva,
Jika RSD<1% =>kurva std
oke