5 XRF Calibration

You might also like

Download as pptx, pdf, or txt
Download as pptx, pdf, or txt
You are on page 1of 38

XRF Calibration

Calibration
Dead Easy, Mate !!
Measured XRF intensity is directly
proportional to analyte concentration,
X-ray
right ?
Intensity
Oops!!

Concentration
Calibration Strategies
• No “best” approach
– most labs have “favourite method”.
– most methods are capable of good results if
used with care.
• Method chosen must meet local criteria of:
– precision
– analysis time
– analysis cost
– contractual requirements (eg ASTM xxx
specified in contract)
Sample Presentation
- spectrometer medium
Medium choice is part of calibration
Sample Presentation
- spectrometer medium
XRF settings will affect intensity
Influence of target
material

• Continuum Influence of target material of the Tube

   1 
I ( )  KiZ   1  2 
  min    
Parameter selection table
Calibration
• In the real world, things are a little more
complicated and a number of corrections are
necessary before the linear relationship
between X-ray intensity and analyte
concentration can be established.
• Fortunately most corrections can be predicted
and the correction factors either measured or
calculated.
• Corrections can be either intensity or
concentration based. There are arguments for
both approaches.
Factors Effecting Calibration
• Background
– the relationship between X-ray intensity and
concentration strictly holds for pure net (ie
background corrected) intensity.
– background is matrix dependent and for some
analytes matrix variations can cause significant
variations in background (eg in coal - can get
factors of 2 or 3 variation in background as ash
level and/or composition of ash varies).
Spectral Overlap

Peak A

Peak B

2
Matrix Effects
• Essentially Two Components
– presence of elements giving rise to spectral
overlap at analyte peak and/or background
measurement positions.
– X-ray absorption effects
• variously termed “matrix”, “inter-element”, “a factor”
and “absorption/enhancement” effects..
Matrix Effects
• X-ray absorption effects
– during XRF analysis all elements present in the
sample will fluoresce and produce characteristic
X-rays.
– if the energy of the characteristic X-rays from any
element is greater than the critical excitation
energy of another element, then there is the
possibility that X-ray absorption will occur.
– the absorbing element may fluoresce under such
circumstances (ie emit secondary X-rays).
Matrix - Absorption Effects
– Consider a sample containing Fe and Ca. During
XRF analysis Fe K and L X-rays will be produced.
– The energy (KeV) of the Fe Ka (6.40) and Fe Kb
(7.06) lines is greater than the Ca critical
excitation energy (4.038).
– Hence some Ca atoms may absorb some of the
Fe K X-rays and may in turn fluoresce.
– The net effect is to reduce the number of Fe K
X-rays and increase the number of Ca X-rays
reaching the detector.
Drift Correction
• All spectrometers will show some degree of
drift with time.
• Need calibration to be time independent.
– Run a drift correction (monitor) sample(s) at
same time as initially run the calibration
samples (Day Zero)
– Run the same drift correction sample(s) at
regular intervals and calculate a drift factor.
– Adjust calibration to compensate for
instrumental drift using the drift factor.
Drift Correction
• Can use one or more drift standards
– may not have all analytes in a single standard.
– can drift correct a single X-ray line using either
one or more standards (typically a high and a
low) standard.
• Suitable drift standards are available
commercially.
– AUSMON – CSIRO Australia
– Breitlanender
Calibration Strategies
• Fusion methods
• Pressed powders
• Metals
• Liquids
• Loose powders
• Environmental filters
Calibration Strategies
• Multi-Point
– Probably most widely used approach.
• “Single Point” (Norrish Approach)
– Approach used many Australian laboratories.
• Semi-Quantitative (“Standardless”)
– UniQuant
– Integral part or option of most manufacturer’s
software packages
Calibration - Multi-Point
• “Original” calibration method.
• Many “local” variations.
• Requires a series of standards of known
composition that cover the required analytical
range.
• Generally 10 or more standards used.
– depends on application, correction method, etc.
• Correlations should be avoided (eg Fe2O3
increasing in direct proportion to SiO2
decreasing).
• The one that most think is best these days
Calibration - Multi-Point
• Relatively quick to set up new calibrations.
• Relatively easy to add additional elements.
– depends on “user friendliness” of software
package.
• Capable of very good results.
• Can use for fusions, pressed powders, metals,
liquids, etc.
“Single” Point
(Norrish Approach)
• Based on a blank and a single synthetic multi-
element standard, plus a series of single
element/oxide beads to determine background
and spectral overlap corrections.
• Multi-element standard approximates
composition of assay samples modified to give
good measurement precision.
• Backgrounds and spectral overlaps all treated
as overlaps.
• Method used in a number of ISO and
Australian Standard X-ray methods (eg ISO Iron
Ore).
“Single” Point
(Norrish Approach)
• Not all software supports this approach (eg
current Siemens/Bruker-AXS and ARL
software does not).
• Time consuming to set up initially.
• Difficult to add additional analytes.
• Capable of very good results.
• No more difficult to maintain than any
other calibration method.
• Applicable to fusions and possibly pressed
powders and liquids.
Calibration Standards
• Requirements for Materials Used for
Calibration Standards.
– Stable.
– Homogeneous
– Known composition and/or high purity.
– Easily weighed (eg not hygroscopic or
deliquescent).
– Preferably readily available and reasonable cost.
Calibration Standards
• What can be used:
– Certified Reference Materials (CRM’s)
• eg NISTS, Eurostandards, SARM, BCS, etc.
– Synthetics (prepared from high purity reagents)
• typically “Specpure” grade (99.99% or better).
– Analysed samples
• eg process samples that have been analysed by a
number of techniques or laboratories.
– Mixtures of the above.
Calibration Standards
- CRM’s
• Certified Reference Materials
– Often only certified for a few elements.
– Analyte ranges often restricted.
– Not all CRM’s are reliable.
– CRM’s for some sample types not available
• eg NZ Steel Melter Slags - high TiO2 and V2O3.
– Pre-treat according to certificate.
Calibration Standards
- Synthetics
• Must use high purity chemicals
• worth doing spot checks - bad batches have occurred.
• Pre-treat to remove moisture and ensure
stoichiometry
o
– V2O5 105 C for 1 hour.
o
– SiO2 1000 C for 15 minutes.
– Cool quickly and weigh. Some pre-treated materials eg
Al2O3 are extremely hygroscopic.
• Can minimise weighing of small masses by making
pre-fused flux reagent mixtures.
Calibration Standards
- Synthetics
• Advantages
– Can design calibration to cover any
compositional range for any analyte.
– Very easy to extend calibration to fit non-routine
samples.
• Disadvantages
– Time consuming to design and prepare.
– Require inventory of (expensive) pure chemicals.
Calibration Standards - Mixtures

• In some instances, there may not be suitable


CRM’s to cover the required concentration or
analyte range. Can overcome by using:
– mixtures of CRM’s
– mixtures of CRM’s and pure chemicals
Calibration
• It is good laboratory practice to prepare at
least some of the calibration standards in
duplicate.
– Easier to identify if a “bad” calibration point is due
to a preparation error or matrix / spectral overlap
problem.
– Improves precision of calibration.
Calibration - Pressed Powders
• CRM’s or well analysed samples are best
approach.
– minimises matrix, grinding and mineralogical
effects.
• Can use mixtures of pure chemicals and/or
CRM’s and pure chemicals, but can have
problems with particle size and other
mineralogical effects.
• Must control particle size of calibration
standards and assay samples.
– Particle size control is absolutely critical for
powder XRF analysis.
Calibration Standards
- Metals
• Only approach available is CRM’s or analysed
samples of similar matrix.
• Sometimes metallurgical history (presumably
related to grain size) can effect calibration.
Calibration Standards
– Loose Powders
• Use same approach as for pressed powders
Calibration Standards
- Liquids
• Multi-point using standard solutions.
• CRM’s and analysed samples
– (eg oils, hydraulic fluids)
• Standard addition
Calibration Standards
- Environmental Filters

• Make up standard filters by depositing an


aliquot of solution onto a filter and allowing to
dry.
– Multi-point using standard solutions.
– Standard addition
• Depositing known mass of pure chemicals as
“dusts” onto filters.
– Requires special apparatus and careful validation.
Matrix Correction
• Can minimise matrix effects by closely matching
calibration standards and assay sample matrices.
• Except in rare cases corrections will still be
required.
• Various correction models available:
– Lachance-Traill
– Rasberry-Heinrich
– Norrish and Hutton
– etc
Calibration Summary

• Background correct
• Spectral overlap correct
• Matrix correct
– the process is normally iterative
– start with an approximate composition
– correct and obtain a better approximation to the
composition.
– repeat until convergence (usually only a couple of
iterations).
Calibration Summary

• Calibration lines should ideally be linear.


• Deviations from linearity may be due to:
– errors in detector dead time correction.
– small errors in the alpha factor determinations.
– alpha factor(s) not constant over entire calibration
range.
• models using higher order terms (eg b and  terms) seek to
theoretically allow for this non-linearity.
• some software packages calculate fundamental alphas for
each sample using the composition of the sample at each
iteration stage (so called “variable alphas”).
Calibration Summary

• ALWAYS validate calibration using CRM’s, cross


analysis with other laboratories or other
techniques.
• Beware of relying just on good correlation
coefficients, figures of merit, etc for the
calibration curve.
– easy to get an r2 goodness of fit index of 0.99998
but the results may be rubbish.
Calibration Summary
• Always back-up calibration data
– Include day one drift correction data.
– Hard copy and electronic
– Ensure calibration data is date stamped
• Allows easy recovery in the event of a problem.
• Requirement of most quality accreditation systems (eg
ISO17025).
• Also back up instrument configuration and
alignment data
– Allows recovery in the event of a system failure.
– Assists engineer during servicing of the instrument.

Link to a good example of how to create a good calibration

You might also like