Download as ppt, pdf, or txt
Download as ppt, pdf, or txt
You are on page 1of 88

Cary Service Training

J. Interlandi
WinUV Software
Cary WinUV software

 Versions
 Version 2.0 (26) currently shipping

 Win UV software is compatible with:


 Windows 95/98 & Windows NT 4.0
 MUST have IE4 installed (Help system)
 Specific PC requirements
 Windows 98: Pentium II
 Windows NT: 64 Meg RAM
 Graphics: 16 bit colors with minimum resolution of 800X600
 CDROM with 16bit sound card and speakers for Help and
Videos

3
Cary WinUV software

 Cary WinUV folder is created on desktop:

4
Benefits of the Cary “Win UV” Software

 Designed with the following Benefits in mind:

Ease-of-Use
Modular design

No Base software


Specific applications for user
requirements

Help available at a click of the mouse


Benefits of the Cary “Win UV” Software

 Designed with the following Benefits in mind:

 Productivity - Multitasking between apps.


e.g., collect data in Kinetics and run another image
of Kinetics for post run calculations

 File Integrity and Security

 Flexibility - Customization of Collects and Report output

 Compatibility - Previous Cary Products PLUS 3rd Party


products
(e.g., Excel)
Common software applications
 Basic installation will consist of
 System Information
 Align
 Simple Reads
 Advanced Reads
 Scanning and Maths
 GLP Admin
 Validation
 Online Help
Individual software applications
 Software applications to target specific markets
 Common software applications
Align, System Information, Simple Reads, Advanced Reads, Scan,
GLP Administration, Validate, Cary Help,
ADL Shell, Cary ADL Help.

 Material Science
Fabric Protection, Sunglasses, Color

 Life Science
Enzyme Kinetics, Kinetics, RNA-DNA, Scanning Kinetics, Thermal

 Pharmaceutical
Dissolution

 Industrial Chemistry
Concentration

8
“Learn how to use
Common Look for All Apps. one app.,
learn how to use
Start/Stop Buttons them all!”

X Value
Y Value
Graphics
Tool Bar

Graphics
Area

Buttons
Area

Reports Area

Hardware Status Sequence Status Results Line


Setup all Method Parameters from ONE window!

 Cary Tab
 Instrument params.

 Accessories Tab
 Analyze Tab
 Setup Calculations

 Reports Tab
 Report format
 Auto Store
 Data Storage options
File System
 The following file types exist:

 Batch files
All method, report parameters, data and calculations
stored in ONE file

 Data files
Data only stored

 Method files
Setup and report parameters stored

 Report files
Generated reports are only stored

 Graphics template
Saves pre-formatted graphics area
File Extension Convention

File Type

* . A BC Application
Type

File Type = A Application Type = BC

if A = B (Batch file) if BC = KN (Kinetics)


M (Method file) CN (Conc)
D (Data file) SR (Simple Reads)
R (Report file) VO (Validation)
SW (Scan)
File Extension Convention

 Application Specific File extensions


eg.
*.BKN (Kinetics Batch file),
*.BCN (Conc. Batch file),
*.BTD (Dissolution Batch file)
*.MKN (Kinetics Method file),
*.MCN (Conc. Method file) etc.

 Only view files specific for the opened application!


The Result is EASE-OF-USE!

13
Help available at a click of the mouse!
 Hint text for every field
Description of fields
Valid ranges

 Interactive help system with “How To’s”


Step by step instructions
align accessories
run different applications
perform typical measures (e.g.. Perform a scan)

 All accessory manuals on-line

 Videos to help with:


replacement of lamp module
alignment of lamp module
alignment of accessories
Flexibility through Customization!

 Customization of Graphics using the tool bar


Add bitmaps to graphics
Add text notes to graphics
Customization of graphics formatting

 Customization of Reports
Publishable quality reports
Ability to add Company logo
Print preview capability
Customize report to obtain all results on ONE page!
Fast Absorbance readings

 F6 (Function key): Rapid Result

Set wavelength (nm)


Blank reading
Sample readings

16
Ease-Of-Use
File Storage Features

 Autostorage of data available


Automatic prompting for file name
All collected data is stored using specified
file name

 Storage on-the-fly with various applications


Dissolution, Kinetics, Thermal, Repetitive
scanning, Concentration

17
Sample Info. Storage
“User Data Form” Concept
 Sample Info. associated
to data
 View info. in Graphics area
(“User Data Form”
Graphics menu option)

 “User Data Form”


facility available

18
Ease-Of-Use

 Designed for Windows 95

Windows 95 conventions used


Drag and drop
Start applications from icons
Use of “explorer” to select files and open applications
Create icons (shortcuts) on the desktop

19
Ease-Of-Use
“Start Methods from icons”
 Start applications from icons

 Icons linked to Method files with shortcuts

 Multiple icons per application

 Icons can be organized into folders on the desktop

20
21
Comprehensive Help available at a
click of the mouse!
 New look interactive help system with “How To’s”

 All accessory manuals on-line

 Print facility produces professionally formatted operation


manuals
Table of contents
Page numbering etc.

 Ability to Print Help as Operation Manual with ‘professional


look’
Viewed section
Whole of Help

22
Comprehensive Help available at a
click of the mouse!
 Hint text for every field
Description of fields
Valid ranges

 Some accessory alignment videos

 Help files written in HTML


Users can modify Help themselves and add hyperlinks
All Help files stored in Varian/carywinuv/Help directory

23
Productivity!

 One “active” program at any time


First opened application controls the instrument

 Multiple applications can be opened allowing you


to perform the following all at the same time:
Data collection
Data analysis, printing and manipulation
Method development
File management

24
File Integrity & Security

 Security
No Method/Data/Report files can be overwritten
when system is running under GLP mode

 GLP app. allows assignment of operator


privileges: eg.
Chemists can create and edit Methods
Technicians can only retrieve and use existing
methods (ie. no method creation/editing
privileges)

25
File Integrity & Security

 Data file Integrity


Automatic storage of data available
Audit logs stored with all data:
Date/Time stamping, instrument serial number, inst.
parameters, operator name, mathematical manipulations all
stored with Data

 Report file Integrity


Modification notification automatically appended to modified
reports
includes time/date stamp, operator name

26
Flexibility through Customization!

 Customize Graphics using the tool bar


Add bitmaps to graphics
Add text notes to graphics
Customization of graphics formatting

 Customization of Reports
Publishable quality reports
Ability to add Company logo
Print preview capability
Customize report to obtain all results on ONE page!

27
Customization of Collects
 User Collect line
 Simple ADL commands
 ‘+’, ‘-’, ‘/’, ‘*’
 factors
 Available in following apps:
 Simple Reads (eg. 260/280 ratio analysis)
 Concentration (eg. 1552 * Read(280))
 Kinetics (eg. 260-320 dual wavelength analysis)
 Provide default set of expressions in drop down combo fields:
 Simple Reads
(Protein estimation expression)

28
Compatibility
 Data Compatibility
 Import
 ASCII XY
 DOS data
 OS/2 data In ONE step!
 Export Win data into Spreadsheet ASCII format
in ONE step!
 Comma delimited ASCII format (*.csv file)
 Compatibility with 3rd party applications

29
Exporting Dissolution Batch File

 Easy to Export data to ASCII *csv format.

 Double clicking on *.csv file in Explorer will automatically open


Excel and load the comma delimited ASCII file.

30
Cut/Paste Report into Word

 Cut/Paste facility
 Graphics, Reports exported quickly into 3rd
party applications (eg. Microsoft Word)

31
System Information
System Information
Company tab

 Company details to be included in the report


 Company name and address
 Company Logo

33
System Information
Instrument tab

 Select instrument models


 Re-initialise instruments
 Enter instrument serial number, this too will be printed on report

34
System Information
Hints tab

 Change hint text default settings

35
System Information
Calibrate tab (Cary 50 only)

 Recalibrate the Cary 50 instrument

36
System Information
Edit tab

 Change User Collect default settings


 enter own pre-defined equations
 set length of drop down list

37
Align
Align

 Ability to align instrument lamps and accessories

 Dip probe coupler


 Fibre optic coupler
 microcell holder
 Transmission accessory etc.

39
Align
Lamps tab

 Change instrument parameters


 lamps, reset lamp hours, save conditions, rescale graph
Cary 50 Cary 100/300 Cary 400/500

40
Align
Graph tab

 Align for maximum signal


 alternative to signal bar
 limited graph handling controls

41
Align
Cary tab

 Change instrument parameters


 wavelength, beam mode, SBW, SAT
Cary 50 Cary 100/300 Cary 400/500

42
Simple Reads
Simple Reads
 Multi-wavelength capability
 Multi-ratio capability
 Raw material QC labs (Pharmaceuticals)
 Ability to change the default User collect equations

44
Simple Reads
Setup tab

 Multi-wavelength capability
 Multi-ratio capability
 Raw material QC labs (Pharmaceuticals)
 Ability to change the default User collect equations

45
Simple Reads
Default parameters

 The following instrument parameters can not be changed

 UVSAT = 1.0 sec


 NIRSAT = 1.0 sec
 UVSBW = 1.5nm
 NIR Energy = 3.0
 Source change = 350nm
 Beam mode = Double
 Grating change = 800nm
 Slit height = Full

46
Advanced Reads
Advanced Reads

 Similar to Simple Reads - but has accessory support

48
Advanced Reads
Cary tab
 Ability to change instrument parameters
 Wavelength, Averaging time, SBW, Y mode, User Collect,
Replicates
Cary 50 Cary 100/300/400/500

49
Advanced Reads
Options tab

 Ability to change instrument parameters


 Source lamps, Beam mode , Detector change
Cary 100/300 Cary 400/500

50
Advanced Reads
Accessories1 tab

 Ability to select accessories and accessory parameters


 Cell changer, Temperature controllers, Temperature probes, RBA.
Cary 50 Cary 100/300/400/500

51
Advanced Reads
Accessories1 tab

 For the 1x1 or 1x0 peltier accessory,


setup the communications settings

52
Advanced Reads
Accessories2 tab tab

 Ability to select accessories and accessory parameters


 SPS-5, RSA control
Cary 50 Cary 100/300/400/500

53
Advanced Reads
Samples tab

 Number of samples, Sample names

54
Advanced Reads
Reports tab

 Setup report format


 Select ASCII conversion for data

55
Advanced Reads
Auto Store tab

 Data storage
 off
 start of run
 end of run

56
Advanced Reads
Reread

 Reread all or selected samples only

57
Advanced Reads
Reread

 Reread, Recalculate

58
Scan
Scan

60
Scan
Cary tab

 Ability to set instrument parameters


 Wavelength scan start and stop, beam mode, SBW, SAT
Cary 50 Cary 100/300/400/500

61
Scan
Options tab

 Ability to set instrument parameters


 SBW, Beam mode, Source lamp, Display options, Signal to noise mode
Cary 100/300 Cary 400

62
Scan
Options tab

 Ability to set instrument parameters


 SBW, Beam mode, Source lamp, Display options, Signal to
noise mode
Cary 500

63
Scan
Independant tab

 Ability to set instrument parameters


 Measurement Mode, UV-Vis and NIR parameters
Cary 500

64
Scan
Baseline tab

 Ability to set baseline ON or recall previously stored


baselines
Cary 50 Cary 100/300/400/500

65
Scan
Accessories1 tab

 Ability to select accessories and accessory parameters


 Cell changer, Temperature controllers, Temperature probes, RBA.
Cary 50 Cary 100/300/400/500

66
Scan
Accessories2 tab

 Ability to select accessories and accessory parameters


 SPS-5, RSA control
Cary 50 Cary 100/300/400/500

67
Scan
Accessories3 tab

 Ability to select accessories and accessory parameters


 Sample Transport, VASRA, Fibre Optic Multiplexer
Cary 100/300 Cary 400/500

68
Scan
Reports tab

 Setup report format


 Select ASCII conversion for data

69
Scan
Auto Store tab

 Data storage
 off
 start of run
 end of run

70
Validate
Validate

 Instrument Performance tests

72
Validate
Tests

 Ability to select specified tests to the instrument


specification

73
Validate
Analyze tab

 Ability to completely automate the validation tests


 Requires the use of the Cell Changer accessory
 Tells you where to put the relevant standards

74
Validate
Reports tab

 Setup report format

75
GLP Administration
GLP Administration

 Default Password ‘Admin’


 Password is case sensitive
 Set new password

77
GLP Administration
Users tab

 Can ADD, MODIFY or DELETE User names


 Bypass GLP checks

78
GLP Administration
Privileges tab

 Can ADD or DELETE specific privileges for different


User names

79
GLP Administration
Groups tab

 Can suspend User names

80
GLP Administration
Applications tab

 Can ADD or DELETE software applications from GLP

81
Help
Help
 Internet Explorer 4.0 MUST be installed
 download from Microsoft web
 Can set Netscape to be your preferred Explorer on your PC

 Advantages:
 more professional looking help
 enhanced index capabilities
 enhanced search engine
 printing Help for hardcopy manual

 Continually use the help system during Cary installations


 If customers know what’s available they don’t
need to call you so often!

83
Help

 pp

84
Help

 Internet

85
Help

 Internet

86
Experiments

 Exp 1: Wavelength Scan


overlay at various scan speeds

 Exp 2: Using Simple reads on Holmium Oxide

 Exp 3: Using simple Reads with User Collect Mode

 Exp 4: Spectral addition of stored files using math

87
Experiments

 Collect single beam scan


Uncorrected baseline

 Collect single beam scan of


D2 emission peak at 656.1nm
Xe emission peak at 541.9nm

 Run Instrument performance tests using


Validation application

 Save the Validation file to desktop


Double Click on this file.

88

You might also like