Seminar On X-Ray Crystallography: Presented by Mounik Rout M.Pharm (PH - Technology) Guided by Dr. Sasmita Kumari Acharjya
Seminar On X-Ray Crystallography: Presented by Mounik Rout M.Pharm (PH - Technology) Guided by Dr. Sasmita Kumari Acharjya
X-RAY CRYSTALLOGRAPHY
Presented By
Mounik Rout
M.Pharm (Ph.technology)
Guided By
Dr. Sasmita Kumari Acharjya
2
INTRODUCTION
X-ray crystallography is a non-destructive technique for determining the molecular structure
of a crystal .
X-ray crystallography uses the principle of X-ray diffraction to analyze the sample .
We can get 3d structure or any sample because it rotates in the sample cell and faces the X-ray
beam in different directions .
DIFFRACTED RADIATION
FLUORESCENCE RADIATION
INCIDENT RADIATION
TRANSMITTED(Absorbed) RADIATION
Sample
Diffracted radiation –It is the most accurate method of analysis with high degree of
specificity than any other X-ray spectroscopy method .
X-ray diffraction(XRD) is the diffraction of incident radiation produced by crystalline sample
according to the atom present in it . 4
X-ray Absorption :-
When incident beam is passed through the sample then some fraction of X-ray photons are
absorbed .
So the no of photons absorbed by the sample is used to measure the concentration of the
sample .
This is similar to any other absorption method like UV-Visible / IR spectroscopy by giving
the information about the absorbing material in the sample .
X-ray Fluorescence :-
When incident beam is passed through the sample then the electrons of the atoms of sample
gets excited by absorbing some energy , when those electrons come to ground state from
exited state they emitt some radiation which have longer wavelength than incident beam .
By measuring the wavelength and intensity of the generated radiation analyst can perform
qualitative as well as quantitative analysis .
5
Schematic diagram of diffraction pattern
6
Production of X-ray
When high velocity of electrons will strike on metal target then X-ray will produce .
It can explain by Bohr’s atomic model .
e¯
e¯
K SHELL
L SHELL e¯
M SHELL
N SHELL
It will release energy in the form of X-ray because of coming from higher orbit to lower
orbit .
The energy of released X-ray will be equal to the difference in the energy between 2 shells .
Atomic nuclei
The striking eˉ interact with the strong electric field of the atomic nuclei constituting the
anode material , which results deacceleration of the striking eˉ and release of energy due to
loosing of kinetic energy and radiation of photons which are responsible for production of X-
rays .
9
10
The process by which photons are emitted by an eˉ is known as ‘Bremsstrahlung’ which
means deacceleration of radiation .
If all initial kinetic energy (Ei) will convert into X-rays ,then velocity of the eˉ become 0 .
This is the condition which will give the X-rays of highest energy and low wavelength .
11
Bragg’s law
X-ray diffraction based on Bragg’s law:-
A C
B D
θ O θ
α
θ
d X θ Parallel planes of crystal surface
Z
θ θ
β d= Inter planner distance
Y
Θ= Glancing angle
C , D = Diffracted X-rays
A ,B = Incident X-rays
When the X-ray is incident onto a crystal surface, its angle of incidence θ, will reflect with
the same angle of scattering, θ. And, when the path difference, d is equal to a whole number,
n, of wavelength, constructive interference will occur.
Bragg’s equation = n λ = 2dsin θ , where n is order of diffraction 12
Derivation of the equation :-
As per the fig. when X-ray falls on the crystal at angle θ then some rays will reflect from upper
plane at same angle θ .
After extrapolating the reflection line then found that there are 2 angle :-
<XOY=θ , <ZOY =θ
And AO=BX , CO=DZ so the path difference
will be XY+YZ ----- 1
Path diff. is defined as an integral multiple of
wavelength =n λ----- 2
So that n λ=XY+YZ---- 3
Taking Sin θ = L L= Length of perpendicular
H H= Hypotenuse
So in ∆XOY , Sin θ= XY
XY=OY×SIN θ OY
=>XY= d × SIN θ----- 4
13
In ∆ZOY , SIN θ= ZY
OY
ZY=OY×SIN θ
SIN θ values = 0 30 45 60 90
ZY= d×SIN θ---- 5 0 ½ 1/√2 √3/2 1
Putting equ.5 in equ.3
n λ= d×SIN θ + d×SIN θ
n λ= 2d×SIN θ ----- 6------ Bragg’s equation
where n = Order of diffraction
λ = wavelength , θ =Glancing angle
d = interplanar distance
Bragg’s equation gives the relationship between:-
1.wavelength of X-rays
2. Interplanar distance in crystal planes
3.Angle of reflection
14
X-ray spectrometer instrument
15
INSTRUMENTATION:-
.
X-ray tube
Collimator Monochromator Sa
m ple
Anode
e¯ Be window
Cathode
Focusing cup Recorder
Amplifier
Tungsten filament
16
Working & Instruments :-
High voltage electric current supplied to heat the tungsten filament to emitt electron from
cathode .
Electrons from cathode striking on anode forms X-ray which will go through Beryllium
window towards the sample through a Collimator reducing undesirable radiation .
Then it passes to Monochromator which gives required X-ray beam to the sample .
Sample have the crystal surface which diffract the X-ray beam to detector made up of
photographic type or counter types .
Detector detects the beam intensity and proceed the signal to amplifier .
Amplifier amplifies the given signal and then the signal proceeds to the recorder screen .
17
X-ray tube :- It is a large vaccum tube containing a heated cathode of Tungsten filament & a Cu
or Mo(Molybdenum) operated at higher voltage up to 60kv .
Collimator :- It passes narrow beam of X-ray by arrangement of 2 closely packed metal plates
separated by a small gap .
Monochromator:- It has 2 type :
1.Filter type – X-ray beam will partially monochromatize passing required radiation by
absorbing undesirable radiation .
2.Crystal type – It is made up of suitable crystalline materials like-NaCl , Quartz , etc.
Detector :- 1. Photographic method – A film is exposed and developed in X-rays passed through
the sample .
2. Counter Method – 1.Geiger Muller counter
2.Propertional Counter
3.Scintillation Detector
4.Solid-state semi-conductor Detector
5.Semi-conductor Detector
18
Detailed image of Bragg’s X-ray spectrometer
19
Bragg’s X-ray spectrometer method
Bragg designed a spectrometer to measure the intensity of x-ray beam following Bragg's
equation i.e. n λ = 2dsin θ
Crystal(sample)
θ
θ
Pb slit
…… … CH3Br
…… …
…… …
G
…… …
…… …
Ionization chamber
20
Working:-
High voltage current is applied on the tungsten filament generating heat which helps the
cathode to emitt electron .
Later the electron strikes on the anode surface producing X-rays which passes through
Beryllium window to the sample.
Collimator placed right after Be window absorbs undesirable radiation and passes required
X-ray beam towards monochromator .
Monochromator converts polychromatic beam to monochromatic beam and then passes it to
sample surface.
The crystalline sample reflect the beam at glancing angle to ionization chamber through lead
slits .
The reflected X-rays ionizes Ch3Br gas to allow current flow in the chamber by electrodes .
Galvanometer present outside to measure the ionization current, which is require to form a
peak in respective to the intensity of X-ray reflected by the crystal .
The ionization current is measured for different values of glancing angle . A graph is drawn
between the glancing angle and ionization current.
21
.
Intensity
θ θ
θ θ
θ θ
θ1 θ2 θ3
Reflection by Crystal Graph
22
.
Crystal
Rotating shaft
Motor
Photographic film will be fixed perpendicular to the incident ray beam inside the cylindrical
camera .
26
POWDER CRYSTAL OR DEBYE METHOD
This technique is a rapid analytical technique used for identification of a crystalline material by
converting it into fine powdered particles .
Cylindrical camera
27
Unwrapped film
Working :-
Powdered sample is taken in a Capillary tube inside the axis of the camera .
The transmitted beam will pass through a hole to minimize the fogging due to direct beam.
On the photographic film diffraction and reflection pattern will be recorded .
Powder particles having different orientation of plane and they will show different pattern of
cone shape whose interaction with photographic film takes place .
From the shape and distance from 2 similar pattern the identification can be done .
And the equation to get this is l= 2πr × Θ or θ = 360l
360º 4πr
28
APPLICATION
Used for the identification of unknown crystalline materials .
For characterization of crystalline materials .
Polymer characterization can be done .
Used to identify impurity .
Determine the unit cell dimensions .
Measurement of sample purity .
Used to carried out drug excipient incompatibility study .
Uses to analyze 3d crystal structure of biological materials .
The powder method is used to determine the value of the lattice parameters accurately .
Now a days often used to probe specific ways in how the structure of a material , drug will
interact in certain environments .
29
1. REFERENCES
2. Critchley, Liam. (2020, October 16). The Applications & Principles of X-
Ray Crystallography.
3. AZoM. Retrieved on March 10, 2022 from
https://www.azom.com/article.aspx?ArticleID=18684.
4. https://youtu.be/HjKI4Kh1RgU
5. Critchley, Liam. 2020. The Applications & Principles of X-Ray
Crystallography. AZoM, viewed 10
6. March 2022, https://www.azom.com/article.aspx?ArticleID=18684.