Oxygen,

Life and Technology


Oxus
M&R / Marketing

1
Contents Ⅰ. M&R

01 Commissioning and inspection.

Ⅱ. Marketing
02 Installation

03 Oxygen Module

04 Air compressor

05

06

2
Ⅰ. Maintenance & Repair

3
1. Commissioning and inspection. Customer Service -OXUS

N
마스터 제목 스타일 편집
1. Basic tools & instruments for commissioning

Q’t
Part Remark Image
o. y

- Battery type
Electric - Cross driver TIP is
1 1
driver
needed

2 Wrench 1 - 250mm

3 Electric wire. 1 - 6mm2/ min.10meter

4 Tester 1 - Ampere /voltage

5 Scissors 1 -for industrial

12/18/2023 4
1. Commissioning and inspection. Customer Service -OXUS

2. Procedure

2. Power Supply ( 6mm2 –wire)

1. O2 tube connection
( Generator – O2 tank)
3 .Switch ON

4. Check the Flow and pressure) 5. Connect tube to the flowmeter outlet port

5
2. Installation ( consideration before installation) Customer Service -OXUS

Temp.&
Humidity

RAK-U06M2E

Ventilation Distance

6
2. Installation Customer Service -OXUS

Installation structure

Oxygen pip-
ing in Hospi-
Pressure tal
sensor

Manual
valve
Open/Closing
valve

Regular Emergency
supply supply

Hospital room

Conventional gas or Liquid oxygen supply


and manifold installation
RAK-U06M2E 7
2. Installation Customer Service -OXUS

 Generally Oxygen Concentrator is installed in the medical gas room.


 Oxygen tube from the oxygen generator is generally connected with main oxygen piping-line after cutting it.

<case1 : Simply connection>


O2 supply

O2 Pipe Line

O2 Manifold

O2 outlet Flowmeter
with humidifies

O2 supply from
Concentrator

< ICU , Operation, Beds >

AC220V ( 6.0 ㎟ )
Cylinders for
emergency
<case 2 : Auto switching valve>

8
2. Installation (References) Customer Service -OXUS

 Air ventilation should be well done , and Oxygen generator should be protected from the rain or hot sun.

AVR

9
2. Installation (References) Customer Service -OXUS

10
3. Oxygen Module change and Test Customer Service -OXUS

1. Procedure – Module Change

1) Disconnect the Black tube connected Air compressor.


2) Disassemble diaphragm valve assembly from the bottom manifold.
3) Disconnect the valve power harness.

<Fig. Unit rear view- Module>


11
3. Oxygen Module change and Test Customer Service -OXUS

2. Test instruments List & Test equipment

No O2 Module
Part Q’ty Remark
.
1 Flow meter 1 Max. 10~15LPM
2 Pressure gauge 1
3 O2 Analyzer 1 - 75%~ , or full range. Pressure gauge
4
Power consumption
1
- Ampere meter compressor O2 Analyzer
meter - Recommending Flow meter
5

< Fig .Test equipment -diagram >

< Fig .Test equipment > < Fig .Test equipment –wall type >
12
4. Compressor parts change and Test Customer Service -OXUS

1. Service Parts ( O2 Unit compressor) & Basic tools.


Item No. Part Q’ty Remark
1 Piston cup 2
2 Cylinder sleeve 2
Parts 3 O-ring 2
4 O-ring (head gasket) 2
5 Valve Flapper 4 - If necessary.
1 T-25 1 - for head
tools
2 T-27 1 - For Piston Cup Retainer
3 Torch 1 - For No.3 screw

< Exploded view & tools >


13
4. Compressor parts change and Test Customer Service -OXUS

2. Procedure – Parts Change

1. Head opening-T-25 2. Valve plate disassembly 3. Unscrew of piston cup retainer

<pic. Damaged parts > < pic. New parts >


14
4. Compressor parts change and Test Customer Service -OXUS

3. Test instruments List & Test equipment

No
Part Q’ty Remark
.
1 Flow meter 1 Max. 180~200LPM
2 Pressure gauge 1
3 Flow restrictor 1
Power consump- - Ampere meter compressor Pressure gauge
4
tion meter
1 - Recommending Flow restrictor
Flow meter
5

< Fig .Test equipment -diagram >

< Fig .Test equipment > < Fig .Test equipment –wall type >
15
5. Facility for M&R Customer Service -OXUS

1. Area ; min. 50 ㎡
1 2

No
Item Q’ty Description
.
1 Work bench 2 - Repair & assmebly
2 Inspection table 1 - Inspection after repair
3 Rack (for parts) 2 - Parts storage
4 Rack (for unit) 2 - Units storage
5 Tool holder 1 - Tool storage
6 Hand cart - Unit transportation

3 4 5 6

16
5. Facility for M&R

17
Ⅱ. Marketing

18
1. Comparison Table (RAK-Series vs PSA plant ( Twin Tower) Customer Service -OXUS

Item RAK-Series (OXUS) PSA plant (General)

Feature

Composition - 3, 6 Oxygen Units - 1 Oxygen Generator


- 2 booster(Pressurize Oxygen pressure up to 6.5 bar) Screw compressor, After cooler, dryer PSA generator, buffer tank etc.

- Continuously Oxygen is supplied - If core parts (compressor or valve) fail, then


Continuity although 1 or 2 Oxygen Unit fail. Oxygen flow is stopped.

- In the unlikely even of ONE module failing,


simply replace the module with a spare.
Maintenance - It is very easy to maintain the Oxygen - Only special technician can repair.
performance - It is very hard to maintain the Oxygen performance.

- Only Oxygen is pressurized up to 5 bar - Inlet Air is pressurized up to 7bar


Power Consumption 2.15KW ( 30 LPM ) / 4.25kW(60LPM) 4KW (40LPM)
- 0.43kW/10LPM 6kW ( 55LPM)  1kW/10LPM

- 230V / 50Hz / 1 phase


- 380V / 50Hz / 3 phase
Power Requirement  Generally, 230V/ 1 phase power is available in
 It is very difficult to provide the 380V / 3phase in local Area.
every local area.

19
1. Comparison Table (RAK-Series vs PSA plant ( Twin Tower) Customer Service -OXUS

Item RAK-Series (OXUS) PSA plant (General)

- Small Space is needed - It needs Air compressor, dryer, tanks and Generator.
Space  It needs only 2.5~3m2 space for installing one machine at - It needs about 4.8m2 space to install 5 parts to set up 1 Generator.
the oxygen room.

50LPM

Length :140cm+40+70+50+50+50(piping)= 4m
Width : 1.2m
Area = 4.8m2
20
2. Comparison Table (RAK-Series vs Oxygen Gas cylinder) Customer Service -OXUS

Item RAK-U06M2E Oxygen Gas cylinder

Feature

- 60LPM (liter per min.)


Production per day - production ; 60liter * (60min./hr)*(24hr/day)
= 86,400 liter

21