Mechanicals

01
Front View
Back View
Reaction Disk
Reaction Disk

Cuvettes
 Cuvette diameter: 5 mm
 Cuvette volume: 1000uL
 150ul minimum reaction volume
 Semi-permanent plastic cuvettes (nine in one set)
 Recommend to replace every three months at least
Reaction Disk

Constant temperature control system


 Solid heating, free-maintenance

T < Tset, Relay on; T >= Tset, Relay off.


Reaction Disk

Encoding sensor

Initial position sensor


Optical System

Optical system

Lamp
 Halogen-tungsten lamp with 2000Hours lifetime

Filter assembly
 Completely sealed system avoid environment interference
 8 wavelengths: 340nm 、 405nm 、 450nm 、 510nm 、
546nm 、 578nm 、 630nm 、 670nm
 Auto-gain adjusting design for accurate reading
Wash Station
Wash Station

4-step wash station

 Wash solution and water


 Real-time dirty cuvettes detection
 High washing efficiency ensures low carry-over

4-step auto washing


 1 step: wash concentrate
 2&3 steps: warmed water
 4 step: Only dry
Wash Station

Initial sensor
Sample/Reagent Disk

Large capacity
 60 reagent positions
 30 R1 reagent positions (Inner ring: No 31 - 60)
 30 R2 reagent positions (Middle ring: No 1 - 30)
 No. 59 for Detergent, No.60 for diluent(DI water)
 60 Sample positions
 Compatible with blood collection tube, minimal tube, sample cup
 For Calibrator, Control and Sample
Sample/Reagent Disk
Sample/Reagent Disk

Cooling system
 24-hour nonstop cooling (With main power switch on)
 Noise-free water cooling design
Sample/Reagent Probe

Probe
 Highly polished design avoid hanging liquid
 Ceramic syringe ensure high precision and free maintenance
 Liquid level detection and tracking
 Collision protection

Precision
 Sample volume: 2~50ul, 0.1ul increment
 Reagent Volume: R1: 150~450ul,R2: 10~300ul, 1ul increment
Sample/Reagent Probe

H_Encoding
sensor

H_Initial
sensor
Sample/Reagent Probe

Anti-collision sensor
Mixer Bar

Mixer
 Highly polished design avoid hanging liquid

Function
 Mix the sample and reagent in reaction cuvettes, improve the
reaction procession
Mixer Bar

Horizontal motor

H_Encoding sensor H_Initial sensor

Vertical sensor

Vertical motor
Mixer Bar
Liquid System

Four-way Four-way
block block
P12

SV02 P3 P4
P1 P2 P5
F5

P11
P10 P6 P7 P8
SY01

Five-way
block F17

Five-way
block
Electrical connection
Power assembly

24V,320W
12V,320W 5V,25W Power
Power Power assembly
assembly assembly

12V,35W
Lamp Power
Analysis Main power
Filter assembly
power switch switch
PCB Boards

Main control
board
Reaction disk
Temperature
control board

Reagent disk
temperature
control board
Main control board

 Parses and converts


commands sent by the
PC;
 Sends command to each
control unit;
 Control motor motion,
fluid path pump and
solenoid valve;
Main control board
Motor
sensor port

Motor
sensor port
Liquid
Liquid pump and
level float valve port
switch port

Communication Stirring
port with PC motor

Serial port for


scanning Liquid level
detection
Temperature photoelectric +24V power +5V, +12V
control board acquisition port supply power supply
Main control board
Step motor ports

Reagent stirring Horizontal motor Cleaning liquid Cleaning needle


horizontal motor for reagent probe syringe pump motor

Reaction disk
motor

Reagent disk
motor
Reagent stirring Vertical motor for Reagent syringe
vertical motor reagent probe pump
Main control board

Power indicator

Indicator Function

DS1 5V power indicator

DS2 +12V power indicator;


DS3
DS4 DS3 3.3V power indicator
DS2
DS4 -15V power indicator
DS1 DS5 +15V power indicator
DS5 DS6
DS6 +24V power indicator
Main control board

Liquid Level Detection Port

Blue: Liquid level signal;


Blown: Probe collision signal;
Black: Power ground;
Red: +12V power supply.
Main control board

Pump/Valve Port JP13


31
Cable Description
1, 2 P1: cuvettes cleaning waste pump 1
3, 4 P2: cuvettes cleaning waste pump 2
5, 6 P3: cuvettes cleaning waste pump 3
1
7, 8 P4: cuvettes cleaning waste pump 4

9, 10 P5: cuvettes cleaning rinse inlet pump

11, 12 P6: cuvettes cleaning purified water inlet pump


13, 14 Lamp relay
15, 16 P7: cuvettes cleaning purified water inlet pump
Main control board

Pump/Valve Port JP13

Cable Description
17, 18 Reserve
19, 20 SV02: Inner wall cleaning valve of sample/reagent probe
21, 22 P11: Outer wall cleaning pump of sample/reagent probe
23, 24 P12: Inner wall cleaning pump of sample/reagent probe

25, 26 P10: Outer wall cleaning pump of mixing rod

27, 28 P8: Coolant water supplying pump


29, 30 Reserve
31, 32 Reserve
Main control board
JP17 Sensor connector
Group 1
Group Function
Pin2 Pin1
1 Horizontal home position signal of stirring rod Pin4 Pin3

2 Horizontal encoder position signal of stirring rod

3 Vertical home position signal of stirring rod

4 Home position signal of reaction disk

5 Encoder position signal of reaction disk Red: Power input terminal


for signal sensor;
6 Home position signal of cleaning needle White: sensor signal terminal;
7 Reserve Green: signal ground;
Black: signal ground;
8 Home position signal of cleaning liquid
Main control board

JP18 Sensor connector

Group Function Group 1

1 Home position signal of reagent disk

2 Encoder position signal of reagent disk

3 Vertical home position signal of reagent probe

4 Horizontal home position signal of reagent probe

5 Vertical home position signal of reagent probe

6 Home position signal of reagent syringe pump


Main control board

JP27 Float Switch Detection Port

Group Function

5, 7, 9, 11, 13, 15
+ 5V Power terminal pins
15
10 Float sensor signal of waste liquid collection
12 Float sensor signal of waste liquid tank
14 1
Float sensor signal of purified water
16 Cleaning liquid float sensor signal
Reagent Temp Ctrl Board

Reagent Temperature Control Board


Reagent Temp Ctrl Board

Radiator fan Cooling pump Scanning window


heater port

+24 V Power
port Thermoelectric
cooler ports

Port for float switch


of cooling water
storage tank

+24 V Power Thermoelectric


port cooler sensor ports

Scanning
window heating
Communication port sensor port
with main control board
Reagent Temp Ctrl Board

DS3 DS8
Indicator Function

DS1 3.3V power indicator

DS2 +5V power indicator


DS2
DS3 +12V power indicator

DS4 +12V circulating pump indicator


DS1 DS4
DS8 +12V radiator fan indicator
Reaction Temp Ctrl Board

Reaction Temperature Control Board


Reaction Temp Ctrl Board

Reaction disk Reaction disk


temperature heating port
sensor port

+12V power +24V power


port port

Communication
port with
main control
board
Reaction Temp Ctrl Board

Indicator Function

DS2 DS1 +12V power indicator

DS2 +5V power indicator

+24V power indicator;


DS3 Indicator light on means
DS3
DS1 the reaction disk is
being heated
LLD Board

Liquid Level Detection Board


LLD Board

Power and Sample/reagent


signal probe port
Liquid detection
connection
indicator
port
DS2

DS1 DS3

Automatic Probe collision Collision


adjustment button indicator sensor
Indicator Function
DS1 +12V power indicator;
DS2 +5V power indicator;
DS3 +9V power indicator;
THANKS