Professional Documents
Culture Documents
Visible and Ultraviolet Spectrophotometry
Visible and Ultraviolet Spectrophotometry
Visible and Ultraviolet Spectrophotometry
Visible andUltraviolet
Ultraviolet
Spectrophotometry
Spectrophotometry
- Interaction of light with atoms and molecules: (Nature of light, various regions
of electromagnetic spectrum, concept of quantization of energy, different types of
energy of molecules).
Absorbance
value
Different molecules will absorb different
wavelengths of light. The wavelengths of light
for UV-visible absorption are from about 200
nanometers to 800 nanometers. This is a section
of the electromagnetic spectrum pictured.
Absorption of light starts with energy of a
certain wavelength in this UV-visible region
being exposed to a molecule.
The light/energy then excites the ground state
(non excited) outer or valence electrons to an
excited state (high energy). The outcome of this
can be measured by a UV-visible
spectrophotometer. The data is shown as a
spectrum with absorption versus wavelength.
This pattern can be used to learn properties of
the molecule. Only certain molecules can absorb
light in this region.
Electronic Transitions:
If light with suitable energy ν hits a molecule in the ground
state ψ1, the energy can be absorbed and the molecule is
raised to an electronically excited state ψ2. The molecule can
return to its ground state via spontaneous emission of a
photon (light).
−− ψ 2 −− ψ 2
↑
Absorption ∆E = E(ψ1) - E(ψ2 ) Emission
= h. ν ↓
−− ψ 1 −− ψ 1
Chromophores: They are molecules that absorb light at these
wavelengths. They are functional groups of a molecule that absorb
light in this UV-Visible region. They are most of the time
characterized by delocalized pi (π)electrons. Pi electrons refer to a
type of bond that occurs between electron orbitals called pi orbitals.
When many of these pi bonds exist in a molecule this allows
electrons to be delocalized or spread out across a molecule. An
example of this type of molecule is pictured below. Many dyes
(colored molecules) are characterized by these delocalized pi
electrons and their color. These molecules can be used for pH
indicators to determine if a solution is acidic or basic. The addition
of acid or base disrupts the delocalized pi electrons. This disruption
causes a color change.
Laws of light absorption
All spectrophotometric methods that measure absorption,
reside upon two basic rules, which combined are known as
the Beer-Lambert law, as follows:
Beer’s Law:
It claims that the amount of light absorbed is proportional to
the number of molecules of the chromophore through which
the light passes.
The absorption of a molecule can be used to determine
the concentration of the molecule in solution. To find
concentration, Beer’s law is used. This is a mathematical
relationship shown by the equation below:
A = abc
A = Absorption, a = absorptivity coefficient, b = pathlength,
c = concentration
The sample container (named a cuvette) is made of a
material that does not absorb light in this region. The
molar absorptivity coefficient can be calculated by
measuring the absorption of a sample of a known
concentration and known pathlength. The absorptivity
coefficient units are dependent on the pathlength and
concentration units. Molar absorptivity units are liters
per mole centimeter (L mol-1 cm-1) when the
concentration units are molarity. The absorptivity is a
measure of how strongly a molecule absorbs light at a
particular wavelength. Once this constant is
determined, unknown concentrations can be
determined from this relationship.
Lambert's law: states that the fraction of light
absorbed by a transparent medium is independent
of the incident light intensity, and each successive
layer of the medium absorbs an equal fraction of
the light passing through it. This leads to an
exponential decay of the light intensity along the
light path in the sample, which can be expressed
mathematically, as follows:
log 10 (I0/I) = kl
where I0 is the intensity of the incident light, I is
the intensity of transmitted light, l is the length of
the light-path in the spectrophotometer cuvette,
and k is a constant for the medium, which is
deciphered by Beer's law.
law
Lambert-Beer Law:
A = log10 (I0/I) = ε × c × l
absorbance ε : molar extinction coefficient,
A: absorbance, coefficient l: path length of sample
cuvette The term Iog10 (I0/I) is called the absorbance (A).
cuvette.
The above equation may be rewritten:
∆A = ∆ ε.c.l
where ∆A is the difference in absorption for the chromophore in the
two environments.
Sometimes the passage of light through the cuvette is described in terms
of transmittance, or transmission (T): T = I/I0 and is generally expressed as a
percentage. It is important to note, however, that only absorbance, not
transmittance, is linearly proportional to the chromophore concentration. In
quantitative analysis, where it is required to obtain the concentration of
substance, therefore, absorbance is more commonly used. The relation between
these two parameters is given by the following:
A = log10(1/T)
Thus, when A = 2 only 1% of the incident light is transmitted, while at A = 3,
only 0.1% is transmitted.
SPECTROPHOTOMETERS (Instruments):
Single-beam Spectrophotometers: are often sufficient for
making quantitative absorption measurements in the UV-Vis
spectral region. The concentration of an analyte in solution
can be determined by measuring the absorbance at a single
wavelength and applying the Beer-Lambert Law.
Single-beam Spectrophotometers are widely used for routine laboratory
measurements at a single wavelength. As shown in Figure 4, only a single light beam
from the monochromator passes through the sample compartment. The absorbance
zero or 100 per cent transmittance is adjusted with a cuvette containing buffer or
solvent. Then the absorbance (or transmittance) value of a sample cuvette containing
sample solution is measured. It is necessary to put the sample and the reference in
the optical beam alternately, by manual operation, typically with an interval of a few
seconds required to change the cuvettes.
Blank
Blank Detector
CH3).
In excited state the s-electrons occupy an anti-bonding
energy level (s*) and the transition is termed s-s*
transition. p-electrons occupy an anti-bonding energy level
(p*) and the transition is termed p-p* transition. While the
n-electrons may occupy s* or p* levels to give n-s* or n-p*
transition. A summary of the different kinds of electronic
levels and transitions is found in Figure 5.
Absorption characteristics of chromophores
1- Ethylenic chromophores
Their bands are difficult to observe in near UV region,
so they are not useful analytically. However,
substitution and certain structural features may cause red
shift rendering the band observable in the near UV region.
Examples;
a) Alkyl substitution; Cause red shift due to hyper-
conjugation and stabilization of excited state.
b) Exocyclic nature, Cause red shift due to relaxation of
strain upon excitation.
c) Attachment to auxochromes, Cause red shift and
increased absorption intensity due to extension of
conjugation.
2- Carbon-heteroatom chromophores
Such as -C=O, -C=N, -C=S, -N=O ….etc. They exhibit some
common characteristics;
n-p* band in the range of 275-300 nm., which is the most
apparent band, has low intensity and long wavelength. This
band undergoes a blue shift on increasing the solvent polarity
due to increasing the energy of transition as a result of
hydrogen bonding
Alkyl substitution; Cause red shift due to hyper-conjugation.
In cyclic ketones an equatorial -halogen results in a small blue
shift (5-10 nm) due to the inductive effect. While axial a-
halogen causes red shift (10-30 nm) due to direct resonance
participation in the excited state, thus lowering its energy.
a) Attachment of heteroatom or group to -C=O e.g. Cl, -NH 2 or -OH
causes blue shift due to resonance elevation of the p * level
hence increase energy of n-p* transition.
3- Conjugated chromophores
Separated chromophores (by two or more single bonds)
have additive effect only because there is little or no
electronic interaction between separated
chromophores, Table 2.
Table 2: Absorption characteristics of selected chromophores
Position of the band, Intensity,
Compound (nm) (liters, cm-1.mol-1)
270 1,450
OH
280 1,430
NH2
269 700
SH
265 240
Cl
261 300
CH3
ANALYTICAL REPORT
Summary (Title, and Aim of the work)
Equipment (Manufacturer, Calibration date)
Material and Analytical Standards (Source)
Chromatographic condition (Column, Flow rate, pH, wavelength,
temperature, mobile system)
Sample preparation
Selectivity / Stability indicating studies
Forced degradation (for stability indicating assay)
Linearity and Range
Limits of detection / quantitation
Repeatability (system suitability),
Accuracy and Recovery
Ruggedness
Robustness
TABLE OF CONTENTS
Summary (Title, and Aim of the work)