Professional Documents
Culture Documents
Instrumentation Chapter 1
Instrumentation Chapter 1
Instrumentation Chapter 1
instrumentation
Chapter one
Introduction
Consulting clinicians on
Test selection
Calibration
How reproducible?
matrix?
• Precision
The closeness of agreement between independent test
results obtained under stipulated conditions
• Accuracy
The closeness of the agreement between the result of a
measurement and a true value of the measurement
Precision Vs accuracy
Selection
Acquisition
Installation
Calibration / Validation
Maintenance
Troubleshooting
Service and repair
Retiring equipment / disposition
Selection Criteria
Use
Performance characteristics
Facility requirements
Cost
Supply of reagents
Ease of operation
Warranty
Service Contracts
Acquiring Equipment
Purchase, Lease, or Rent
Donor provided
Contract considerations:
Parts Manual
Installation
Operators’ Manual
Trial period
Contents of service contracts
Installation Checklist
• Prior to installation:
– verify physical requirements have been met
Safety checks, electrical, space, ventilation, water supply,
ambient temperature, etc.
– confirm responsibility for installation
• Upon receipt:
– verify package contents
– do not attempt to use prior to proper installation
• If required, ensure the equipment is installed by the
manufacturer
Installation
• After installation
– Establish inventory record
– Define conditions
– Develop and implement protocols for calibration,
performance verification, and operating procedures
– Establish maintenance program
– Provide training for all operators
Calibration of instruments
During installation
After maintenance
After displacement of instrument
After systemic error
Periodic calibration
Conditions of use
Usage rate
Skill level of personnel
Degree of accuracy expected
Costs of calibration
Instrument maintenance
1. Preventive maintenance
Schedule Maintenance before the breakdown of the
instrument, the schedule can be
Daily
Weekly
Monthly
This type of maintenance is performed by laboratory
professionals
2. Curative maintenance
Unscheduled Maintenance after the breakdown of the
instrument and this type of maintenance is
performed by trained biomedical engineer
Maintenance Program
Assign responsibility
Oversight of all laboratory equipment
Individual responsibilities
Develop written policies and procedures
Train staff
Keep records
For each piece of equipment:
Establish routine maintenance plan
Establish required function checks
Develop a list of spare parts
Troubleshooting
• Safety
• Fewer interruptions of work
• Lower repair costs
• Elimination of premature replacement
• Less standby equipment
• Identification of high maintenance cost
• Reduction of variation in test results
• Greater confidence in the reliability of results
Retiring Equipment / Disposition
• When?
– When experts indicate not repairable
• Why?
– Hazardous
• How?
To remove dirt, chemicals such as ions and substances so that only pure H2O
remains for laboratory uses.
– Type II
• Qualitative urinalysis, rinsing laboratory ware, qualitative reagents or
stains
– Type III
• Tap water for washing laboratory ware
25
Criteria established to determine levels of purity
26
Four Main Methods of Water Purification
• Distillation
• Ion Exchange
• Reverse Osmosis
• UV Oxidation
27
Distillation
28
Distillation
29
Ion Exchange Filtration
• Principle:
– some or all ions are removed with a membrane ion exchange filter
• Good secondary water purification step
• Produces Type I water if start with distilled water and then filtered
with anion and then cation filters
30
Ion Exchange Filtration
Disadvantages:
• Relatively expensive process especially if two-column system
• Ion exchange columns need replacement periodically
31
Reverse Osmosis Filtration
• Principle:
– pressure forces water through semi-permeable membrane to
remove many organic and inorganic chemicals.
• Dissolved gases remain.
• Type I water is produced from reverse osmosis filtration if first start
with distilled water
Disadvantages:
• Very expensive process
• Reverse osmosis filter columns need replacement periodically
32
UV Oxidation
• Principle:
– Water exposed to UV wavelengths and ozone will sterilize bacteria
but not remove them.
• Type I sterilized water is produced after membrane purification
followed by UV oxidation process.
Disadvantages:
• Very expensive process
• Not commonly available in developing countries
• Filter columns and UV lamp need replacement periodically
33
Sterilized type I water
34
Laboratory Sterilization
Autoclave
Basic Definitions
36
Purpose
Methods
– Moist heat (autoclave)
– Dry heat
– Chemical sterilisation
– Filtration
38
Moist Heat Steriliser/ Autoclave
39
Temperature
• Autoclaving at 1210C for 15-20 min. is required for the fastest route.
Time
• 1210C for 15-20 min. is required for the fastest route of sterilization but a lower
temperature and longer sterilization time can be used.
• It takes certain period of time for the saturated steam to penetrate the entire
load and for heat transfer to occur. Time for heat up needs to be added to
total time
41
Autoclave Safety
42
Methods of monitoring the performance of Autoclave
Daily
– Follow manufacturer’s instructions and monitor time,
temperature and pressure gauges
– Simple screen with Sterilization strip quarterly or annually
– Spore strips or Indicator chemicals
43
Other Less Used Sterilization Methods
47
Class I Safety cabinets
• It is an open-fronted work chamber which is exhaust-ventilated to
provide personnel and environmental protection only by means of an
inward air flow away from the operator
• 100% air exhaust
• The exhaust air being filtered through a HEPA filter before being
discharged from the cabinet.
• Class I cabinets should be used with agents of low to moderate risk.
48
Class I Safety cabinets
Class II Safety Cabinets
• A Class II biological safety cabinet is a partially open-fronted work
chamber which provides protection for personnel, the surrounding
and/or experiment against contamination by means of HEPA filtered air
flowing in a downward uniform and unidirectional (laminar) manner.
• The means of creating a “barrier air flow” at the work opening is not
only by means of an inward air flow from the operator but also from
the inside to create air curtain.
• The downward air velocity should be sufficient to afford protection
against cross-contamination and prevent outflow.
• There are three types of class II cabinet
50
Type A1 cabinet
• The Type A1 cabinet has a minimum inflow velocity of 75 ft./min.
• The filtered makeup air is divided equally over the work surface at
about two to six inches above the work surface.
• Exhaust is drawn at the bottom of the cabinet where it rises to the
top.
• At the top of the cabinet, 70% of the air recirculates through the
supply HEPA filter, the other 30% of air exhausted through the
exhaust HEPA filter.
Type A1 cabinet
Type A 2
• Has a minimum inflow velocity of 100 ft/min.
• A negative air pressure plenum surrounds all contaminated plenums
that are under positive pressure
Type A 2
Type B1
• The Type B1 cabinets have a minimum inflow velocity of 100 ft/min.
• These cabinets must be hard‐ducted to an exhaust system( In contrast
to the type A1 and A2 cabinets, 70% of air from the rear grille is
exhausted and only 30% is recirculated).
Type B2
• Type B2 cabinets have a minimum inflow velocity of 100 ft/min
• Type B2 cabinet is expensive to operate because no air is recirculated
within.
• this type is mainly found in such applications as toxicology laboratories.
• cabinets of these types generally monitor the exhaust flow, shutting off
the supply blower and sounding an alarm if the exhaust flow is
insufficient.
• Used in contaminated environment
Class III cabinets
• It is a totally enclosed and gas-tight structure.
• Work procedures in the cabinet are carried out through replaceable
arm-length gloved sleeves.
• The cabinet is supplied with air through a HEPA filter and exhausted
through two HEPA filters mounted in series.
• The cabinet is operated at negative pressure and provides total
protection for use with all categories of biological agent.
• The negative pressures within a Class III cabinet should be a minimum
of 200 pa(pascal).
• The air flow should also be sufficient to allow for two of the glove ports
to be open and achieve a minimum velocity of 0.75m/s through the 59
Class III cabinets
Safety cabinet effectiveness
• The place of safety cabinets are important. The main problems are
caused by current of air from doors and windows and the movement
of people and therefore, safety cabinets should not be sited near
doors.
• HEPA filters are highly efficient in removing viable microorganisms
with a quoted efficiency rate is 99.997%.
• The achievement of barrier air flow should be confirmed by
contamination challenge tests.
• When successful containment is reached, exhaust velocity should
then be measured and recorded
62
Decontamination of safety cabinets