Professional Documents
Culture Documents
Training Presentation SW70 - Advanced
Training Presentation SW70 - Advanced
Training Presentation SW70 - Advanced
15.06.2021
Device Front:
1. On/Off button
2. Display
3. Knob
4. Soft Keys
5. Charge LED
Micro SD card
socket
Ventilation:
IP67 protected
The Basic Mode contains minimal menu entries for inexperienced users to operate the device with minimal
knowledge and can be fully customizable in Expert Mode.
The device always starts in Basic Mode when the license is installed (installed by default)
Changing the modes is done via menu selection „Expert Mode” or ”Basic Mode” in the main menu window.
Initially Expert Mode is password protected with the default password (SONOTEC)
Password can be changed or disabled (no password set) Device Settings> „Set Expert Mode Password”.
While in Expert mode all changes done will correspond to the Basic Mode after switching the mode or device
restart.
To customize the menu for Basic Mode there is a special menu entry in Device Settings> „Setup Basic Mode
Menu”.
In case of password recovery need contact SONOTEC to recieve a special file.
cancellation
Password input and
confirmation
Password check
right
false
Expert Mode
There is a new menu item in the „Device Settings”: „Setup Basic Mode Menu”.
The deactivated menu items are not displayed in the basic mode (empty box).
[Press the rotary knob]: The selection screen should appear (just like when transferring the files) with all
possible menu items (main and submenus arranged hierarchically).
+ Test Preparation
Visible in
Long press
Basic Mode
Not visible in
Basic Mode
Short click
Select/deselect
If the menu item has a submenu, a "+" (submenu closed) or "-" (submenu opened) is displayed in
the box in front of it. If not, there is no submenu.
Short knob press to select and deselect the corresponding menu item; if menu item with submenu is
selected/deselected (+), all submenu items are activated or deactivated.
By pressing and holding the knob, submenu opens/closes.
Basic Mode is preconfigured on delivery (to have access to all functions start Expert Mode).
When leaving the window, by pressing SoftKey “OK” the data is automatically saved.
The selection does not affect the expert mode which always have all options on.
Loading setups
Data from the loaded setup overwrites the current settings in Basic Mode (even if it was created in
expert mode and the functions are not active in basic mode).
Following Test Options turned Off for Basic Mode will not be loaded from the setup files. For
example: Data Logger, Corrosion Module, Multi-layer measuerments when disabled, the setup that
contains those options will not be loaded.
When loading setups, all the parameters that are not available in Basic Mode would be changed
automatically based on setup parameters.
The Basic Mode is license controlled software option which is installed in all devices for no extra
cost.
If this license is not installed or deleted, the device always starts without password in expert mode.
Basic mode is also considered for skilled distributors who wants to set up the devices for their
customer’s applications and help them in daily life to easy operate the device.
Backwall Echo
Attenuation off
Gate 1 Gain
Measured ampl.
indicator
Gate 2 A – Scan presentation
Measuring
mode Lower thickness limit
Measuring values
Long press
AutoRange: open the softkey „Measurement Range”, long press and hold pressed the rotary knob (> 2s) for
activating or deactivating AutoRange
Long press
AutoGain: open the softkey „Gain”(Test > Receiver), long press and hold pressed
the rotary knob (> 2s). For activating or deactivating AutoGain or
Press the rotary knob and rotate while pressing left or right for activating or deactivating the AutoGain.
Non soft buttons have to be open.
Note: This operation has to be executed fast before menu pops out.
OR
Press & Rotate
at the same
time <0.5s
Status Display: The state of AutoGate, AutoRange and AutoGain will be displayed
in the status line at the top right of the screen.
All automatic functions are great when they work properly. There is no possibility to create the automatic
algorithms which are going to work on every possible testing scenario. Therefore additional parameters
controlling those functions were introduced:
Auto Gate Dead Range – so called blanking. Sets the time (starting range) from which Auto Gate will start
looking for echoes. Everything which is before will not be detected by Auto Gate. This can be used to
prevent the Auto Gate to detect the interference signals especially when high gain is used (probe crosstalk,
interface echo, noise, etc.). To adjust it go to: Test Preparation \ Automatic Functions \ Auto Gate Setup
The zone in which Auto Gate is excluded from echo detection is marked by a red mesh and it is
dynamically presented at the time of change on the A-Scan. Echoes which are within the red mesh will not
be detected.
Dead Range set to 3.13us which
Dead Range set to covers almost 10mm
1.4 us which
Visible red mesh on the right side
covers at current
of the A-Scan and the first echo
sound velocity
from 6mm inside which was
approximately
excluded from measurements.
4.5mm.
Auto Gate detected 2nd end 3rd
Visible red mesh echo instead of 1st and 2nd.
on the right side of
This technique can be used to
the A-Scan
improve the reading on low
thicknesses outside the focus
point, where first echos are not
very well separated.
Auto Gain Range – this is a parameter which controls the Auto Gain dynamic range. In many applications
there is a need to limit the Auto Gain dynamics due to the many imperfections (probe, test piece material,
surface quality, etc.).
This parameter when set corresponds to the +- value of the gain range which will be added or deducted to
the base gain. For example, if the base gain is set to 50dB and Auto Gain Range is set to 40dB it means
that the whole dynamic range of the gain control is from 10dB-90dB.
Base gain can be set freely for each probe/application by turning the Auto Gain Off, adjusting the gain to the
desired value and switching Auto Gain On again. Auto Gain will start from the set gain.
It is important to set as much base gain so the amplitude will cross the first gate threshold, from that point it
can adjust itself automatically if the dynamic range doesn’t limit it too much.
When amplitude is too low to trigger the Auto Gate it can be manually added by the knob.
When the base gain is set to high the echoes will overflow the full screen height and would not be detected
correctly. Instead of those echoes some transverse mode conversions would be detected.
Auto Gain finish the adjustment when measured echo is at 60% FSH, in case of E-E mode when the lower
echo from two gates is on 60% FSH.
It is important to optimize the base gain and the Gain dynamics for each application.
Auto Gain Range can be accessed from:
Test Preparation \ Automatic Functions \ Auto Gain Setup – SoftKey1: Auto Gain Range
+- dB value which
is added or
deducted from
base gain
In order to load the probe from database go to: Test Preparation \ Probe
and press the knob to open the database.
From the list select the desired probe and press the knob to activate it.
To add the probe, go to: Test Preparation\ Probe and press the knob to open the database.
When database is opened, press SoftKey1 “New” to introduce the new probe to database.
NOTE: Depends on the chosen mode (Basic or Expert) the probe parameters window will contain full or
limited parameters which are described in next slides:
Set the Parameters as requested and press SoftKey “OK” to save the probe in database.
Expert probe parameter window is often too complicated for less expierienced operators. For simplicity, the
following changes are introduced:
Variant I – Expert mode Variant II – Expert mode Variant III – Basic mode
Zero In Air Zero On Block Zero On Block, In Air analogic to V1
Some of the parameters which can be adjusted to improve the echo response (obtain an echo from the actual
measurement piece and try changing following parameters):
a. Transmitter:
- Voltage: can be set from 50-400V and it is directly connected with the energy transmitted to the probe.
The more we set the the higher should be the echo response unless we reach the probe or transmitter limit. The
point to set would be when echo is at maximum height and doesn’t go up anymore.
NOTE: some high frequency probes available on the market works only with limited voltage. When that probe would
be connected to the device which is set to 400V it might irreversibly damage the probe. Always check the maximum
probe parameter from the datasheet and device setting to avoid such a situations. All SONOTEC’s probes can work
with 400V voltage.
b. Damping:
- There are two settings for the damping of the transmitter. 50 or 400Ohm. It needs to be checked in the
probe datasheet what is the damping of the probe and set the closest impedance in the device. Switching it
when echo is visible will cause amplitude to change. Set the optimum damping considering your application.
Echo height vs. echo width preference.
c. Pulse width:
- By changing the pulse width we are changing the transmitting impulse duration time. Changing that time
directly influence the crystal excitation and therefore directly affecting the amplitude response from a
test object. Every single probe will have slightly different window for optimal settings. Adjusting it might bring few dB
change in signal height. Setting too much will cause the amplitude to drop and even to create more
unwanted mode converted signals.
Waveform averaging – waveform averaging is a mathematical function that computes the average of a
sequence of waveforms. This is useful for removing random noise from a signal.
There are several settings for averaging:
1x meaning that averaging is OFF
2x - software computes 2 consecutive images and remove the noise
4x - software computes 4 consecutive images and remove the noise
8x - software computes 8 consecutive images and remove the noise
Signal averaging causes measuring rate to drop as software is computing “x” number of A-Scans images.
The higher the averaging multiplier the bigger the noise reduction and bigger measurement rate drop.
Very useful feature when high gain is used which causes higher random noise.
Important Note:
In our calibration the name should be
„TS5i 4-8mm”
It reflects the industry need of easy data collection, management and reporting,
Due to selectable 1D, 2D and 3D structure and microgrids it is dedicated to virtually all
thickness surveys,
Rich data model allows an operator to report all thickness relevant settings for each
measurement point,
Easy reporting with pre configured, linked and structured Excel file (*.xlsx)
Access to the raw data for integration with external management software (*.db file)
Alarm based color coded cells: green – within a limit, red – outside the limit
Comments databases for easy comment assignment (to points, calibration or file)
NOTE:
Old values are only visible when the probe is
coupled to a test piece and reading is obtained.
Have you ever had a problem to corelate your measurement point on your asset with a
point on the grid?
Together with database import/export functionality all grid descriptions could be edited
on PC and loaded in SW70.
Q: Would you like to edit your database on PC before inspection and open it in SW70?
A: Not a problem anymore!
SW70 doesn´t create all cells at the time of file creation but adds points when they are collected (for
memory saving reasons)
That means that when the file is created it doesn`t contain any measurement cells.
To create a full empty grid there is a special function in device menu: „Prefill
Measurement Points“
Only empty (without any measurements) grids can be populated with this function
This function automatically creates all the cells which then are ready for editing
SONOTEC recommends: DB Browser for SQLite – 3rd party software for editing
Dual element probes are consisting of two piezoelectric crystals separated by acoustic barier – transmitting
and receiving one. In order to increase the probe sensitivity, and focus it to a certain depth, elements are
placed inside the probe in certain angle – roof angle. This creates an effect of V shape ultrasonic beam which
propagates in the material. This V shape beam propagates not exactly paralel to the surface therfore an error
is introduced to the measurement. Mainly in low thicknesses.
Two point calibration in relatively short range compensates this error good enough but the problem araise
when for some reasons it is not possible to do two point calibration.
V-Path Error Correction extends the accuracy of the measuerment by reducing the
error occured due to roof angle of the dual element probes.
It is not meant to work with single element probes as the beam shape of this probe is
parallel to the surface.
To assure the proper use of this function calibration has to be done before performing
it.
NOTE: It is important to set the correct sound velocity of the test block which is used to
perform V-Path correction function.
4. Place the probe on first step of the block (in our case
2.50 mm)
Click „New” to create the file Enter the Name and confirm „OK”
Ultrasonic Solutions I Made in Germany
Test Options: Time encoded B-Scan
Multi - Layer test option is a special mode of measuring thickness where the operator is able to measure thickness of
each layer separately and the total thickness is calculated. Materials which consist of up to 4 different layers
(difference in sound velocity) can be measured precisely.
If we con-sider a material which has at least two different layers, in normal thickness mode only one sound velocity
would be used to calculate the thickness and the error will occur. For example paint (2mm) on steel (10mm): paint
usually has less than 2500m/s velocity and steel has 5940m/s – factor 2,4. Using sound velocity of steel for that
measurement (5940m/s) going through the paint instead of 2mm we measure 4,8mm (error in paint) plus 10mm of
steel, so total 14,8mm instead of 12mm.
It is important to mention that for the Multi-Layer measurement device have to be able to display the signal from each
layer with a nice quality and separation, which is not that easy.
Proper probe and device settings have to be used.
The Layers have to be significantly different in acoustic impedance so the reflection may occur. If not, the amount of
reflected signal will not be enough to proper measure it. If the acoustic impedance is very big it will cause almost total
reflection from the first layer and measurement will not be possible too.
NOTE: Multi Layer module works only in full manual mode so all the automatic functions are disabled when started.
There is a need to switch them back on manually after closing the module in order to use them.
NOTE: Gates which are used in that mode are different than normal measuerment gates. They are called Lx and can
be controlled only from Multi-Layer module menu.
NOTE: Gate L2 can‘t overlap or be set before L1 to prevent proper layer measurements.
NOTE: Proper calibration of probe delay has to be done before measurement
1. Place the probe on a test object and recognize the echoes from each layer.
2. Place L1 gate over first layer backwall amplitude.
3. Use Gate Selection Soft Key to switch to second gate – L2
4. Place L2 Gate on second layer backwall reflection. If you have more layers continue the same with Gates 3 and 4.
5. The measuerment for each layer is displayed as L1 and L2 value and the total sum at the bottom.
In next slide you can find the measurements using TXS7,5i probe made on 8mm steel block with different coating
thicknesses (aproximately 1,2mm and 3,3mm).
Sound velocity of the material changes with significant change of temperature. In elevated temperature sound
propagates faster and when it gets low it is slower. It is important to calibrate and measure the material of the same
temperature to be accurate. In industrial environment it is very difficult to maintain this requirement because if the
specimen is hot, the calibration block should also be heated to the same temperature. If this is not a case the error
occurs. In order to take more precise measurements in such a situations, temperature compensation of the sound
velocity is needed. This software function compensates the temperature changes between the calibration piece and
measured element.
1. Using this function operator can calibrate on the block with room temperature (or any other).
2. Go to the Temperature compensation function and enter calibration and measurement temperatures.
NOTE: Temperatures has to be measured by external device.
NOTE: Standard probes have temperature working range from -20 to +60 Celsius degrees. If the surface temperature
exceed this range, a high temperature probe is needed.
3. Set the Temperature coeficient (default coeficient of 181.8 ppm/K is meant for steel and longitudinal wave only)
NOTE: The coefficient used as default (181.8 ppm/K ) allows to compensate the temperature only in steel by using
longitudinal wave probes (normal single or dual element probes). For angle beam probes the cooeficient is going to
be different, same for other materials than steel.
4. Turn the compensation On and start taking compensated measurements.
Auto Hold is an standard software test mode option which saves the last reading on the screen after the probe is
uncoupled.
It helps the operator to read the last reading for example if the equipment is put back and the reading was not
remembered (often useful when readings are written on a pice of paper).
This last reading is displayed as a hollow font and it stays there until next proper measurement is taken.
To activate the function go to: Test Preparation \ Test Mode – SoftKey2: Auto Hold (On \ Off)
Before start make sure the measuerment mode is set to peak for angle beam probes. To verify
the state check bottom line of the screen:
To Change it to peak go to: Test > Measuerment Mode: [SoftKey] Meas. Mode change from Flank to Peak
1. Prior starting the calibration go to: Test Preparation > Calibration > Two – Point Calibration
2. When it is selected two SoftKeys appears: „Ref. Position 1” and „Ref. Position 2”
3. As we have to set 25 and 100mm as reference points use those buttons to set it
4. When it is finished couple the probe to a test block and see if you have two backwall echoes
visible on your A – Scan, like below:
Important Note:
1. Make sure probe is connected and
coupled.
2. Increase the gain by rotating the wheel
3. If echoes are still not visible please use
SoftKey „Measuerment Range” to extend
it to cover two echoes
1. To start the calibration go to: Test Preparation \ Calibration \ Two – Point Calibration and press the wheel to
execute it.
2. Couple the probe to V2 block and maximize the echo.
3. Hold the probe in that position and try not to move it or
use freeze function which can be set to top button.
4. Use „G1 Start” SoftKey to move the blue gate on first
echo
5. When properly gated push „Next” SoftKey
After succesful calibration sizing curve is often needed to estimate the indication size.
DGS is a mathematical, evaluation method of the flaw size which can be crated based on refference back wall
echo from the standard block (K1 or K2).
1. To draw DGS curve go to: Test preparation > Evaluation Curves > DGS > Set DGS Reference
Note: It is recommended to use K1 block 100mm radius as a refference point of DGS as the radius
reflection factors are already set in the probes data for SONOTEC probes. For other manufacturers probes
factors should be similar, if not amplitude correction has to be adjusted (explanation on further slides).
5. Couple the probe to the test block and maximize the echo reflection from any radius of the calibration block.
6. When amplitude is on maximum select it by gate by „G1 Start” Soft Key and press „Set Reference”
7. When succesfull the curve will be drawn on the screen, similar to one below.
Note: created curve corresponds to the back wall reflector size
(full reflection). In order to make it usefull for flaw size evaluation
(comparison) we need to set correct DSR size for our inspection.
For example if we need to inspect anything based on DSR size
3mm or 1,5mm we should go back one step in the menu.
[Test Preparation > Evaluation Curves> DGS > (Soft Key 3) DSR
Size and set that size to the size we will verify the curve.
After the change the curve should look like on the left picture.
DGS veryfication can be made on the side drilled hole of K1/V1 calibration block (in our case 3mm)
After succesful calibration sizing curve is often needed to estimate the indication size. Creation of a DAC curve
requires a special prepared test block of the same material and SDH size which require testing procedure.
That means, if we are looking for flaws bigger than 3mm in steel we should have a steel block with 3mm SDH
on different depths (similar to weld thickness) in order to create precise DAC curve.
1. To draw DAC curve go to: Test preparation > Evaluation Curves > DAC > New DAC Curve
2. Couple the probe to DAC block and maximize the echo
from first reflector.
3. Use „G1 Start” SoftKey to move the blue gate on that
echo
4. Press SoftKey „Add Point”
After succesful setting the DAC curve there is a need to adjust it to the measurment conditions by adding the
gain (for example 6dB).
The preffered way to do so is:
1. Use the transfer correction to lower the curve (+6dB correction) (Test Preparation > Evaluation Curves >
2. Add 6dB gain using the wheel the ammount of gain which is needed by normal gain (the DAC curve will go
up as well)
In order to save the A-Scan pattern start the device in Flaw Detector mode.
Go to: Test\ A-Scan options and when selected use Soft Key1: „A-Scan Mode“.
Switch the mode to „Comparison“ so the Soft Key4 appear: „Comparison Scan“.
Couple the probe to the test piece and press the „Comparison Scan“ when ready.
Note: When the gain, measuring range and other important parameter changes the pattern disappears
and needs to be created/recalled.
BEA – is a flaw detection technique applied in ultrasonics to identify or confirm the presence of defects by
observing a reduction in the amplitude of the echo reflected from the backwall. A typical example is in the
inspection of forgings using a 0° probe. The amplitude of the echo from the backwall will remain approximately
constant in clean material. The presence of defects between the probe and the backwall reduce the amount of
sound reaching the backwall and, consequently, reduce the size of the backwall echo.
Planar laminar defects are likely to produce direct pulse-echo responses, which can be detected as signals
observed in the range between the probe and the backwall echo. The BEA provides confirmation of the presence
of such defects. Clusters of small defects, such as porosity, may not produce a reflected signal of sufficiently
high amplitude to be detected by direct pulse-echo. However, depending on the number and density, they may
scatter the sound sufficiently to be detected by BEA.
For an inspection of a forging, the flaw detector gain, required to detect the signals from potential defects, will be
sufficiently high such that the backwall signal will be significantly greater than full-screen height. The BEA
function allows a separate lower gain to be applied to a gate towards the end of the timebase, so that the
backwall echo can be fully observed on the screen during the inspection and any changes in amplitude can be
monitored.
Functionality:
There is a special 3rd gate available to perform that function “BEA Gate”.
BEA gate allows the operator to apply an independent gain level to the A-Scan within the BEA gate width.
It allows the operator to reduce the amplitude of the backwall echo when general gain level is high (often
used to find small flaws).
The high general gain makes internal reflectors easy to see and the BEA gain setting allows the operator to
observe the peak of the backwall which in normal case would be above full screen height.
Backwall variation could be the result of flaws that scatter sound energy or reflect sound energy away from
the transducer.
Suppressing the backwall signal can also enhance detection of defects close to the backwall echo by
allowing visibility of the defect and backwall peaks even when the two signals overlap at the baseline.
Operation:
To activate/deactivate Backwall Echo Attenuation (BEA) go to “Test Preparation – Test Options – Backwall
Echo Atenuation” and use soft button “BEA Enable” to switch it to “ON”/”OFF”.
BEA gate appears on the screen and it can be controlled by the soft buttons on the bottom of the screen.
BEA gate operates only in manual mode
As BEA is an additional gate, there are 2 normal gates available for usual testing scenarios