Professional Documents
Culture Documents
3 - Calibration
3 - Calibration
• Measuring Element
Sensors, Transmitters, Transducers, Process Switches
• Receiving Element
Indicators, Recorders, Controllers, Alarm Units, Totalizers, Computer-based
instruments & systems
Response time is how quickly an element reacts to a change in the measured variable or
produces a 100% change in the output signal due to a 100% change in the input signal.
For example, the response time of a temperature sensor determines how quickly it indicates or
records a change in temperature.
Dynamic error is the difference between a changing value and the momentary instrument
reading or the controller action.
Hysteresis is a property of physical
systems that do not react immediately
to the forces applied to them or do not
return completely to their original
state. Systems that exhibit hysteresis
are systems whose condition depends
on their immediate history. Frictional
or magnetic forces may cause
hysteresis. Hysteresis affects valve
actuators by slowing down the
response to a changing control signal
because of a worn linkage or Hysteresis is the property of a control element that
overtightened packing nut. results in different performance when a measurement
is increasing than when the measurement is decreasing.
Linearity is the closeness to
which multiple measurements
approximate a straight line on
a graph. Linearity is usually
measured as nonlinearity and
expressed as linearity.
Nonlinearity is the degree to
which multiple measurements
do not approximate a straight
line on a graph. Hysteresis is the property of a control element that
results in different performance when a measurement
is increasing than when the measurement is decreasing.
What is Calibration?
1. A set of operations wherein known values of a
quantity are applied to an instrument and
corresponding readings or output values are
recorded under specified conditions.
2. A comparison of the reading of a measuring
instrument against the reading of a standard
instrument of higher accuracy.,
• It may include adjusting the instrument to
read correctly (not in all cases).
• During calibration, an instrument is checked
at several points throughout the range of the
instrument.
Calibration according to Legal
Metrology http://legacy.senate.gov.ph/lisdata/2973226440!.pdf
(R.A. 9236 of 2003)
Calibration is an operation that, under specified
conditions, in a first step, establishes a relation
between the quantity values with measurement
uncertainties provided by measurement standards
and corresponding indications with associated
measurement uncertainties and, in a second step,
uses this information to establish a relation for
obtaining a measurement result from an indication.
INSTRUMENT CALIBRATION BLOCK DIAGRAM
c b
a b–a= Note: Instrument
c Sensitivity must
be adjusted until
c=0.
Note: IMS and OMS are commonly known as CALIBRATORS.
Typical Instrument Calibration Errors
with Linear Response
Recall that the slope-intercept form of a linear
equation describes the response of any linear
instrument:
Example 1:
A flow transmitter is ranged 0 to 350 gallons per
minute, 4-20mA output, direct responding. Calculate the
current signal value at a flow rate of 204 GPM.
Example 2:
A pneumatic temperature transmitter I ranged 50 to
140 degrees Fahrenheit and has a 3-15 PSI output signal.
Calculate the pneumatic output pressure if the temperature is
79 degrees Fahrenheit.
Example 3:
A pH transmitter has a calibrated range of 4pH to
10pH, with a 4-20mA output signal. Calculate the pH sensed
by the transmitter if its output is 11.3mA.
Instrument Errors:
An "error" is defined as the algebraic difference between the
instrument indication and the actual value of the measured
variable. Typical errors that are corrected by performing a
calibration include:
∙ Zero error
∙ Span error
∙ Linearity error
Some instruments are provided with a means of adjusting the
zero and span of the instrument.
Calibration helps to correct such errors and improve the
accuracy of the instrument's readings.
INSTRUMENT ERRORS
Zero Error - produces a parallel shift of the input-output curve
This usually require multiple adjustments of the “zero” and “span” screws while
alternately applying 0% and 100% input range values to check for
correspondence at both ends of the linear function
INSTRUMENT ERRORS
Linearity Error - produces non-linear input-output response curve
(Input Standard)
(UUT)
+ -
D/P Transmitter
(Range: 0-200 mBar) Atm.
(Output Standard)
PRESSURE CALIBRATOR
Calibration range:
A portion or whole part of the instrument range the
instrument is calibrated; expressed by stating the lower and
upper range values.
"Zero" value:
The lower end of the range.
Span:
The algebraic difference between the upper and lower
range values.
Tolerance is the
maximum deviation that is
accepted in the design of the
user for its manufactured
product or components.
Calculated
INPUT VALUES OUTPUT VALUES (mA) Error in
Accuracy in
mA
% of Span
% INPUT INPUT EXPECTED ACTUAL
Calculated
INPUT VALUES OUTPUT VALUES (mA) Error in
Accuracy in
mA
% of Span
% INPUT INPUT EXPECTED ACTUAL
∙ Digital Multimeters
∙ Signal Simulators
∙ Distance/height measuring devices
∙ Plotting/Recording devices
∙ Indicating devices
“Material Type” Calibration Standard
It means:
• All measuring instruments must be calibrated.
• All standards used to calibrate the measuring
instrument must also be calibrated.
• The calibration standard must, in turn, be
calibrated.
∙Numeric Value
∙Unit of Measure
∙Indication of Accuracy
THREE TYPES OF ERROR
(Input Standard)
(UUT)
12 mA
+ -
D/P Transmitter
(Range: 0-200 mBar) Atm.
(Output Standard)
PRESSURE CALIBRATOR
Measured Value= 12 mA + Uncertainty of
TYPICAL BENCH CALIBRATION SET-UP
Measurement
Uncertainty
Uncertainty is a parameter associated with the result of
a measurement that characterizes the dispersion of the
values that could reasonably be attributed to the
measurand.
Uncertainty analysis is required for calibration labs
conforming to ISO 17025 requirements to evaluate
and identify factors that affect the calibration accuracy
of the equipment and process instrument.
Uncertainty
Calibration technicians should be aware of basic
uncertainty analysis factors, such as environmental
effects and how to combine multiple calibration
equipment accuracies to arrive at a single
calibration equipment accuracy. The combined
accuracy is determined by calculating the square
root of the sum of the squares.
The two ways to estimate uncertainties
Expanded Uncertainty
(U) = uc k
= 0.29 (2)
= 0.58%
Where:
uc = Combine standard uncertainty
K = Coverage Factor
Level of Confidence:
Most of expanded uncertainty calculations
are based on coverage factor (k=2) and
confidence level of 95% (1 chance in 20 that
the value of the measurand lies outside the
interval).
Some other coverage factors are:
k = 1 for a confidence level of approximately 68 percent
k = 2.58 for a confidence level of 99 percent
k = 3 for a confidence level of 99.7 percent
4. Compliance to ISO-17025:2000
Technical Requirements
∙ Human Factors (Personnel)
∙ Environmental Conditions
∙ Test & Calibration Methods and Method Validation
∙ Test/Calibration Equipment
∙ Traceability
∙ Handling of test and calibration items.
ISO/IEC 17025
0% 100 deg C 0% 4 mA
12𝑚𝐴
8 𝑚𝐴
% INPUT
10-point Calibration Procedure:
A calibration procedure which utilizes a 10 input and
output test points. This is the most widely used calibration
procedure.
Test points commonly used are 0, 25, 50, 75, 100, 75, 50,
25 and 0% of the input and output span.
12𝑚𝐴
8 𝑚𝐴
% INPUT
Hysteresis
Hysteresis calibration error is when an
instrument reacts differently to increasing
and decreasing inputs.
While performing such a directional calibration test, it is important not to overshoot any of the
test points. If you do happen to overshoot a test point in setting up one of the input conditions
for the instrument, simply “back up” the test stimulus and re-approach the test point from the
same direction as before. Unless each test point’s value is approached from the proper
direction, the data cannot be used to determine hysteresis/deadband error.
Non-Conventional or Specific Calibration
Procedure:
Actual UUT
Expected Error in % of
Input (%) Input (mbar) Indication in Error (mA)
Output (mA) Span
mA
Actual UUT
Expected Error in % of
Input (%) Input (mbar) Indication in Error (mA)
Output (mA) Span
mA
As left
0 0 4 4 0 0
Data 25 25 8 8 0 0
50 50 12 12 0 0
Table 75 75 16 16 0 0
100 100 20 20 0 0
Finalize Calibration Certificate.