MPAES Accessories Applications
MPAES Accessories Applications
Applications,
Accessories,
Software
and
Food and Agriculture Environmental
Questions
• Software Overview
• Improved performance
• Matrix handing to match
Chemical / flame AA
Petrochem • Low running costs
• Ease of use
• Runs on air
• Improved performance
• Improved Geochemistry/
• No flammable gases
Environmental performance
• Runs on air
Metals • Unattended operation
(non –regulated • Moderate pricing
methods) • Unattended
operation
12
Inert OneNeb nebulizer
10
Conventional concentric nebulizer
Proportion
8
of Total
0
0 10 20 30 40 50 60 70 80 90 100
Parameter Setting
Nebulizer Glass concentric
Spray chamber Single-pass glass cyclonic
Read Time 3s
Replicates 3
Sample uptake delay 10 s
Stabilization time 5s
Fast Pump On (80 rpm)
Background correction Auto
Nebulizer Pressure 140 to 240 kPa (auto optimized)
Improved productivity
• Automate analysis using SPS 3 autosampler
• Fast pump during uptake and rinse
6
MP - AAS correlation
y = 0.9918x - 0.0009
5
MP - ICP correlation
y = 0.974x + 0.0027
Au MP-AES (ppm)
4
MP - AAS
MP - ICP
Correl=1
3 Linear (Correl=1)
Linear (MP - ICP)
Linear (MP - AA)
2
0
0 1 2 3 4 5 6
Au ICP/AAS (ppm)
MP - normal AA range
MP - Extended Range
FA QC
FA Q
3 MP ~ 267 Rational
AA ~ 242 Rational
0
0 1 2 3 4 5 6
100
80
60
MP ~ Cu 324.754 Rational
MP ~ Cu 327.395 Rational
MP ~ Ag 328.068 Rational
40 MP ~ Ag 338.289 Rational
AA ~ Cu 327.395 Rational
AA ~ Ag 328.068 Rational
20
0
1 3 5 7 9 11 13 15 17 19 21 23 25 27 29 31 33 35 37 39 41 43 45 47 49
-20
40
35
30
25
MP ~ Cu 327.395 Rational
20
MP ~ Cu 324.754 Rational
AA ~ Cu 327.395 Rational
15
10
0
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23
“With the MP-AES instrument Agilent seems to have hit a homerun for both gold analysis and base metal
analysis. It provides superior detection limits and vastly expanded calibration range – this enables quick easy
analysis at trace levels as well as eliminates the need for time consuming dilutions of higher grade samples. All
of this at an affordable price of well below that of ICP and only slightly higher than top of the line AA units. ”
- Bobby Joe Reichel
Newmont Mining Corporation
Other Comments –
Very low detection limits for gold compared to AA and graphite furnace
Extended calibration range for Cu & Ag (Cu-MP up to 125 ppm vs. Cu-AA up to 35 ppm; Ag-MP up to 50 ppm vs. Ag-AA up to 10
ppm.
Purge
• O2 in the atmosphere absorbs low UV (<185nm)
• EGCM
Carbon build up in
injector
• OneNeb nebulizer
Nebulizer
blockage
Parameter Setting
Nebulizer Inert OneNeb
Spray chamber Double-pass glass cyclonic
Sample tubing Orange/green solvent resistant
Waste tubing Blue/blue solvent resistant
Read Time 3s
Replicates 3
Stabilization time 15 s
Rinse time 45 s
Fast Pump (80 rpm) On
Background correction Auto
Pump speed 5 rpm
Great
recoveries
for all key
elements
Great recoveries
for all key
elements
1.5
1.4
1.3
1.2
1.1
1
Normalized Conc
0.9
0.8
0.7
0.6
0.5
0.4
0.3
0.2
0.1
0
0:00 1:00 2:00 3:00 4:00 5:00 6:00 7:00 8:00
Time (h:mm)
Parameter Setting
Nebulizer Inert OneNeb
Spray chamber Double-pass glass cyclonic
Sample tubing Orange/green solvent resistant
Waste tubing Blue/blue solvent resistant
Read Time 3s
Replicates 3
Stabilization time 15 s
Fast Pump (80 rpm) On
Background correction Auto
Pump speed 5 rpm
Parameter Setting
Nebulizer Inert OneNeb
Spray chamber IsoMist cooled spray chamber (-10 oC)
Sample tubing Orange/green solvent resistant
Waste tubing Blue/blue solvent resistant
Read Time 3s
Replicates 3
Stabilization time 30 s
Fast Pump (80 rpm) On
Background correction Off-peak
Pump speed 5 rpm
Essential minerals
Parameter Setting
Nebulizer Glass concentric
Spray chamber Single pass glass cyclonic
Sample tubing White/white
Waste tubing Blue/blue
Read Time 10 s
Replicates 3
Stabilization time 10 s
Fast Pump On (80 rpm)
Pump speed 12 rpm
Nebulizer pressure 160 – 180 kpa (auto optimized)
Background Correction Auto
NIES No. 10c Rice Flour High Purity Stds CRM Wheat Flour
NIST 1577 Bovine Liver High Purity Stds CRM Oyster Tissue
“The results for our aqua regia soil extract samples are convincing,
and those results are in very good agreement with the round robin,
which was performed Europe wide on those samples.”
- Frank Symossek
Sachsenforst
“We are using MP-AES instrument for last couple of months, and we
found instrument is best to detect trace metals compare to AAS. In
AAS sensitivity for B and P is very poor and fuel cost is very high,
whereas using the MP-AES, sensitivity is very, very good and fuel cost
is very low. In AAS, with elements like Al, Ca, Mo,Ti, V, Si etc, it
consumes high quantities of Acetylene and Nitrous Oxide gas, and is
very tricky - whereas using the MP-AES, it is very easy to detect.
There is also no need to change Hollow Cathode Lamps which is time
consuming and costly.”
- S.S. Mhatre
CERA Laboratory