02 BeneVision N17N15N12 Service Training V2.0 en

You might also like

Download as pptx, pdf, or txt
Download as pptx, pdf, or txt
You are on page 1of 42

BeneVision N17/15/12 service training

PMLS Global Technical Support Department


Version: 2.0
No.:MJH-17002-BeneVision N17N15N12
Service@mindray.com
Contents
System Principle

PCBA Introduction

Software Upgrade

Hardware Upgrade

Troubleshooting
Contents

System Principle

PCBA Introduction

Software Upgrade

Hardware Upgrade

Troubleshooting
System Principle

Model No. Factory Code S/N


N17 6610 F2-
N17 ICU* 6610 F3-
N17 OR* 6610 F4-
N15 6620 F5-
N15 ICU* 6620 F6-
N15 OR* 6620 F7-
N12 6630 F8-
N12C 6631 F9-

* Note: ICU and OR models will be released later.


System Principle CMS iPM Beneview

- System Overview Remote


display 2
Remote
display 1

View others

LAN

Hospital Wired Infrared ear


network temp modulee
printer CIS
service
Quickly temp module Telemetry
AP

WiFi
AC power
Patient
monitor
AC power cord
USB USB Adaptor
Remote
U disk control
IR DP & USB
DIAP
UART RS422
USB USB
RS422 IR
anesthesia USB Display(with
touch
screen)
benelink
Parameter
module

mouse keyboard
scanner UART SMR
IR
Parameter
TDS
module
System Principle
- Block Diagram
System Principle
- Main unit Structure
Rear
Housing

Front
Housing

Bottom
Housing
System Principle

N12/N15/17 construction

1. AC-DC module
2. Recorder (optional)
3. Main board
4. iView assembly (optional)
5. Inner module rack COM board
6. MPAN module (optional)
7. Alarm lamp board
8. Touch screen control board
9. Power switch button board
10.Battery interface board
11. Speaker

Note: Front shell part 7/8/9/11, others are part of back shell.
Contents
System Principle

PCBA Introduction

Software Upgrade

Hardware Upgrade

Troubleshooting
PCBA Introduction Troubleshooting

- Power Supply
PCBA Introduction
- N17/15 Main Control Board

Full Configuration (N17) Standard Configuration (N17/15)

Note: Standard configuration board cannot support iView and independent display modules.
PCBA Introduction
- N12 Main Control Board

Full Standard
configuration configuration

Can not
support DVI
PCBA Introduction
- N17 Independent display board

 One DVI port


PCBA Introduction
- iView System

Top view Bottom view


PCBA Introduction
- iView System

ComE module

SSD card

Cell battery
iView Carrier Board
PCBA Introduction
- 4-slot Infrared COM Board (N12)

Multi-function port
(for TDS)

BNC connector

Standard configuration Full configuration


PCBA Introduction
- 6-slot Infrared COM Board (N17/15)

Multi-function port
(for TDS /SMR)

BNC connector

Standard Configuration (N15) Full Configuration (N17/15)


PCBA Introduction

Power Switch Board

Alarm Lamp Board


PCBA Introduction

MPAN board
WiFi module

Note : MPAN and WiFi functions are optional.


PCBA Introduction

N12 power board N17/15 power board

Note : The power supply of N12 is 63W, but N17 and N15 are 100W.
PCBA Introduction

Front housing assembly - SHARP Front housing assembly - NLT

- The front housing assembly includes display, touch screen, front housing, touch screen control
board and cable.
PCBA Introduction
Battery :

115-018012-00 Li-ion Bat Pack (11.1V4500mAh LI23S002A)


051-002723-00 Battery Interface Board PCBA

N12 N15/17

Note: N12 battery location is different from N15/17.


Contents
System Principle

PCBA Introduction

Software Upgrade

Hardware Upgrade

Troubleshooting
Software Upgrade

Method 1: Network upgrade

Tool :

 System Update Tool


 Upgrade Tool --'Merak.tool'
 System upgrade package -- 'System Software'
 Network cable , HUB
 Keyboard(optional)
 PC
Software Upgrade

Method 1: Network upgrade

Steps :
1. The IP address of the monitor must be set to 77.77.1.xx, and the subnet mask is
255.255.255.0

2. Choose the 'Merak.tool' model package, and select the product type with 'Merak'

3. Enter the upgrade model: Continue to touch the touch screen or press both 'F4'and 'F5'
in the keyboard when you press the power switch button.
Software Upgrade
Method 1: U disk upgrade

Tool :

 U disk
 Boot --'Merak_Installer.pkg'
 System upgrade package -- 'System Software'
 keyboard(optional)

How to make U disk?

1. Created a directory in the U disk: UPGRADE_AMP/Merak;


2. Put the boot software 'Merak_Installer.pkg' into the directory 'UPGRADE_AMP/Merak'
3. Put the upgrade software (pkg /mpkg) into the directory 'UPGRADE_AMP/Merak'

Note: U disk Format requirements: FAT format(2GB or above ,such as Kingston)


Software Upgrade

Method 1: U disk upgrade

Upgrade steps :

1. Inset the U disk to the monitor.


Main board has 4 USB connectors, you
can inset it into any one connector.

2. Enter the upgrade model.


Continue to touch the touch screen or
press both 'F2'and 'F3' in the keyboard
when you press the power button.
Contents
System Principle

PCBA Introduction

Software Upgrade

Hardware Upgrade

Troubleshooting
Hardware Upgrade
Items Part No. Descriptions Notes
WiFi 115-044521-00 2.4G/5G WiFi module upgrade package
MPAN 115-044522-00 MPAN upgrade package
Embedded For N17/N15/N12
115-044522-00 Recorder upgrade package
Recorder
1. For N12 only
N12 Full 2. Replace main
configuration 115-044508-00 N12 full configuration upgrade package board and Infrared
COM Board
115-044553-00 iView upgrade package ( no main board/WIN7 )
iView 115-044578-00 iView upgrade package ( with main board/WIN7 )
115-050002-00 iView upgrade package ( no main board Win 10 )
Independent display upgrade package ( no main
115-044554-00 board)
Independent
For N17 only
display Independent display upgrade package ( with main
115-044579-00
board )
Contents
System Principle

PCBA Introduction

Software Upgrade

Hardware Upgrade

Troubleshooting
Troubleshooting

Password
 Maintenance user: 888888
 Maintenance factory: 332888
 Configuration: 315666
 Demo: 2088
Troubleshooting
- How to check Software version ?

1. Select 'Main menu'→select 'System


>>'→select 'Version >>'.In the display
menu, you can check the system software
version.

2. Select 'Main menu'→select 'System


>>'→select 'Maintenance>>' enter the
required password → select 'Version'. In
the display menu, you can check the
version information of the system
software and modules.
Troubleshooting
- How to export log?

1. Insert U disk in the USB port of patient


monitor.

2. Select 'Main menu‘ →select 'System


>>'→select ‘Maintenance' →enter factory
password(332888) →select ‘Factory
Maintenance’ →select ‘Monitor
information’ →select ‘Export log’.
Troubleshooting
- How to export configuration and user maintenance settings?

1. Insert U disk in the USB port of patient


monitor.

2. Select 'Main menu‘ →select


‘Configuration >>'→select ‘Manage'
→enter password(315666) →select ‘Save
current setting’

3. And then, select ‘Export Configuration’


→select ‘User maintenance settings’
and the current configuration →select
‘Export’
Troubleshooting
- How to export real-time patient data?

1. Installing real-time export data


tool

2. Setting local IP address

3. Chose the parameters what


you want to export.

Note:
• Please learn more information from the instruction of ‘The Manual for Monitor Data Export
Tool’
• If you want to use this function, the N series patient monitor system software should be V02.03.00
or above
• The format of the export data is CSV and can read with Excel.
Troubleshooting
Power On/Off Failures
Symptoms Possible Reasons
AC mains not connected or insufficient battery power

Connector cable defective


Power on Power switch and led board defective
failure
Power module defective
Main board failure
Power supply protection
 MPAN Failure
Symptoms Possible Reasons
Antenna not installed
MPAN module failure

MPAN Failure The cable connecting MPAN module and internal module rack COM board is defection
Internal module rack COM board is defective
MPAN module software version incorrect
Troubleshooting
Display Failures
Symptoms Possible Reasons

Connect cable defective

Blank screen or cannot LCD display defective


properly display ,
but the patient monitor still Main board software damaged
operates normally Backlight drive damaged
Display drive damaged
Connect cable defective

Touch screen damaged


Touch screen no response
Main board software damaged

Main board defective


Troubleshooting
Module Rack Failures
Symptoms Possible Reasons
Parameter module software different between SMR software
SMR power supply defective
SMR communication board defective
SMR can not identify
the parameter module Interior cable defective
Module rack backplane board is defective
The software of the Module rack backplane board is defective
Main board defective
Parameter module failure
Cable defective or improperly connected
Internal module rack
can not identify the Parameter module software different between module rack software
parameter module
Internal module rack power supply defective
Main board defective