Professional Documents
Culture Documents
Chapter 6-UV-Vis Spectros
Chapter 6-UV-Vis Spectros
Source
Monochromator
Slit
• Principle
Detector
• Beer’s Lamberts Law
• Operation of UV-VIS Spectroscopic
• Typical Industrial Application
ELECTROMAGNETIC
RADIATION
SPECTROSCOPY
The study of the interaction of light with matter
The study of interaction of electromagnetic radiation
with matter
• Energy (light or electromagnetic )applied to
matter can be absorbed, emitted, cause a
chemical change, or be transmitted
• The main application of UV-VIS Spectroscopy,
which depends on transitions between electronic
energy levels, is in identifying conjugated pi
electron systems
Electromagnetic RADIATION
• Light is a form of electromagnetic radiation.
WAVE
Frequency (v):
- It is defined as the number of times electrical field radiation oscillates
in one second
- The unit for frequency is Hertz (Hz)
1 Hz = 1 cycle per second
Wavelength:
- It is the distance between two nearest parts of the wave in the same
phase i.e. distance between two nearest crest or troughs.
Electromagnetic Radiation
• The relationship between wavelength and frequency can be
written as:
• c=
• = wavelength in centimeters (cm)
• c = velocity of light (3 x 1010 cm/s)
• = frequency in reciprocal seconds
(s-1)
E = h
• The higher the frequency, the greater the energy of the wave.
• The shorter the wavelength, the greater the energy of the wave.
Interaction of EMR with
matter
Electronic Energy Levels:
• At room temperature the molecules are in the
lowest energy level Eo
• When the molecules absorb UV-visible light from
EMR, one of the outermost bond/lone pair electron
is promoted to higher energy state such as
E1,E2, ..En, etc is called as electronic transition and
the difference is as
E=hv=En-Eo
Where (n=1,2,3,…etc)
Electromagnetic Spectrum
12
Electromagnetic Spectrum
13
Absorbance and Complimentary Colour
Absorption
Absorption
• The process by which energy is transferred
from an electronic wave to an atom or molecule
and causes it to move from ground state to an
excited state.
Energy is emitted by
electrons returning to
lower energy levels
3rd
Excited 2nd
States
1st
Energy is absorbed as
electrons jump to
higher energy levels
Ground State
19
Absorbance VS
Transmittance
Absorbance VS Transmittance
The absorbance is a logarithm function of
transmittance,
A = - log T
Note:
Since transmittance is often expressed as percentage
and called percent transmittance (%T= T x 100 %), A
also can calculate as,
A= log100- log %T
A = 2 – log %T
Note:
1. Absorbance is zero when transmittance is
100%, ie. No light absorbed by the sample
P = Po
b = 1 cm
T = 1.0 or 100 % and A = 0
P0 P
Transmitted light = P
Incident light = Po
0.09
0.08
0.07
Coffee Sample
0.06
Abs A
0.05
0.04
0.03
0.02
0.01
0.00
250 300 350 400 450 500 550 600 650 700 750 800 850
NM
100
98
96
94
%T 92
90
%T
88
86
84
82
80
250 300 350 400 450 500 550 600 650 700 750 800 850
NM
BEER
LAMBERT
LAW
CONCENTRATION of a
substance in solution is
directly proportional to the
ABSORBANCE of the
solution
Beer’s Lambert Law
states the relationship between the absorbance of a
solution and the concentration of the absorbing species.
A=abc (1)
Where a = Proportional equation; absorptivity (Lg-1cm-1),
b = optical pathlength (cm)
c = Concentration (g L-1).
A=bc (2)
Where A = absorbance,
= molar absorptivity (L mol-1cm-1),
b = optical pathlength (cm)
c = Concentration (mol L-1).
It is also known as absorption law or beer-Lambert law
• This Law tell us quantitatively how the amount
of attenuation depends on the concentration of
the absorbing molecules and the pathlength
over which absorption occur.
• Absorption may be presented as:
Transmittance ( T = P / Po )
or Absorbance ( A = log Po / P)
Absorption of Radiation
Using the Beer's Law
Beer’s law expressed in equation (1) and (2) , can be used
in several different way:
1. Calculate molar absortivity of species if their
concentrations are known.
2. Can use to measure value of absorbance to obtain
concentration if absorptivity and pathlength are
known.
3. More often , we use a series of standards solution of
the analytes (different concentrations) to construct a
calibration curve, or working curve of A versus c. Then
it can be used to determine the concentration of
unknown.
Calibration Graph
Limitations of Beer-Lambert
Law
• deviations in absorptivity coefficients at high concentrations
(>0.01M) due to electrostatic interactions between molecules in
close proximity
• scattering of light due to particulates in the sample
• fluoresecence or phosphorescence of the sample
• changes in refractive index at high analyte concentration
• shifts in chemical equilibria as a function of concentration
• non-monochromatic radiation, deviations can be minimized by
using a relatively flat part of the absorption spectrum such as the
maximum of an absorption band
Experiment: Analysis of Food
Colour
Objectives:
Chemicals
100 ppm Carmoisine stock (100 ml), Unknown concentration
of Carmoisine sample (two), Distilled water
Apparatus
Perkin – Elmer UV/Vis Spectrophotometer Lambda EZ210,
Sample cuvettes, path length 1 cm, Volumetric flask 50 ml
(five) , Pipette 5 ml, 10 ml and 25 ml (one each),Rubber bulb
(three),Beaker 100 ml (one),Graduated cylinder 50 ml
(one),Dropper (one),Labeling sticker,Tissue paper
Procedure
1) Prepare a serial dilutions (5 ppm, 15 ppm, 25 ppm, 35 ppm
and 45 ppm) from the 100 ppm food dye stock.
2) Calculate the volume needed, V1 from the 100 ppm food
dye stock for all dilutions by using dilution formula
(M1V1=M2V2).
3) After preparing the serial dilutions, Fill a cuvette with 45
ppm dilution and another cuvette with blank solution, insert
them in the sample compartment. Wipe clean the sides of the
cuvettes.
4) Record the absorbance readings and look at the standard
calibration graph produced.
5) Determine the concentrations of Unknown #1 and
Unknown #2.
Standard Solution
Result
No Sample Carmoisine Concentration Absorbance (nm)
(ppm)
1 Standard 1 5 0.214
2 Standard 2 15 0.591
3 Standard 3 25 1.025
4 Standard 4 35 1.333
5 Standard 5 45 1.714
6 Unknown 1 ??? 0.775
mol L-1)
= 1.75 x 10-7
PRACTICE
MAKE
PERFECT!!!
1. A sample in a 1.0 cm cell is determined with
spectrometer to transmit 80% light at a certain
wavelength. If the absorptivity of this substance at this
wavelength is 2.0 Lg-1cm-1, what is the concentration of
the substance? Ans: 0.048 g/l
Typical UV VIS
spectrophotometer consists of a
(1) SOURCE, a
(2) MONOCHROMATOR,
(3) SAMPLE CELL,
(4) DETECTOR and
(5) READ OUT
Instrumentation and Basic Components
All spectrophotometers
require a SOURCE of
continuous radiation
over the wavelengths of
interest.
Sources of UV radiation
Sources used in UV-Vis Spectrophotometers are continuous sources.
• Continuous sources emit radiation of all wavelengths within the
spectral region for which they are to be used.
• Sources of radiation should also be stable and of high intensity.
• The combined output of these two bulbs is focused on to a diffraction
grating
Continuous
Sources
Visible and
Ultraviolet
near IR
radiation
radiation
Tungsten Deuterium
Lamp Lamp
320-2500 nm 200-400 nm
REQUIREMENTS OF AN IDEAL SOURCE
• It should be stable and should not allow
fluctuations
• It should emit light of continuous spectrum of high
and uniform intensity over the entire wavelength
region in which it’s used
• It should provide incident light of sufficient
intensity for the transmitted energy to be detected
at the end of optic path.
• It should not show fatigue on continued use.
Source of visible radiation
• The tungsten filament lamp is commonly employed as a source of
visible light.
• This type of lamp is used in the wavelength range of 350 - 2500 nm.
• Tungsten/halogen lamps contain a small amount of iodine in a quartz
"envelope" which also contains the tungsten filament.
• Tungsten/halogen lamps are very efficient, and their output extends
well into the ultra-violet. They are used in many modern
spectrophotometers
(2) monochromators
All spectrophotometers
require a MONOCHROMATORS
for selecting a narrow band
of wavelengths from the
source spectrum.
Monochromators
• It is a device used to isolate the radiation of the
desired wavelength from wavelength of the
continuous spectra.
• All monochromator contain the following
component parts:
1) An entrance slit
2) A collimating lens
3) A dispersing device (a prism or a grating)
4) A focusing lens
5) An exit slit
• Polychromatic radiation (radiation
of more than one wavelength)
enters the monochromator
through the entrance slit.
• The beam is collimated, and then
strikes the dispersing element at
an angle.
• The beam is split into its
component wavelengths by the
grating or prism.
• By moving the dispersing element
or the exit slit, radiation of only a
particular wavelength leaves the
monochromator through the exit
slit.
(3) Sample cells
All spectrophotometers
require a SAMPLE CELL for
holding the sample (usually a
solution) and a reference cells
Cells
• One of the two divided beams is passed
through the sample solution and second
beam is passed through the reference
solution.
• Both sample and reference solution are
contained in the cells.
Sample Containers
Sample Container, usually called cells or cuvettes,
must be transparent to the radiation which will pass
through them.
Types of material normally used are:
1. Quartz or fused silica required for the UV region
(wavelength less than 350 nm) and may used in
the visible region and out to about 3000 nm in the
IR region.
2. Silicate Glass ordinary used for the region of 375
to 2000nm because of its low cost compare quartz
3. Plastic cells also used in visible region sample
Cleaning of Cell
• Use the solvent in which the substance was dissolved
• After cleaning rinse with distilled water
• Never use hydrofluoric acid for cleaning – will etch the
surface
• Never use ultrasonic cleaning – may cause cavitation
(4) detectors
All spectrophotometers
require a DETECTORS for
converting radiant energy
into electrical energy.
Detector
Usually uses photo-electric devices. It converts the
radiant energy (hv) into an electrical signal.
Examples of detectors are:
Photomultiplier tube
Linear photodiode array
Charge –coupled devices
Detector
The photomultiplier tube
• The photomultiplier tube is a commonly used detector in
UV-Vis spectroscopy.
• Photomultipliers are very sensitive to UV and visible
radiation. They have fast response times.
• Intense light damages photomultipliers; they are limited to
measuring low power radiation.
Double beam spectrophotometer
detector
• Generally two photocells serve the purpose of
detector in UV spectroscopy.
• One of the photocell receives the beam from
sample cell and second detector receives the
beam from the reference.
• The intensity of the radiation from the reference
cell is stronger than the beam of sample cell.
• This results in the generation of pulsating or
alternating currents in the photocells.
(5) Read out
All spectrophotometers
require a DEVICE TO READ OUT
the response of the detector.
Types of Instrument
05/21/2024 91
(b) Double Beam UV – VIS
Spectrophotometer
Has two light paths, one for the sample and one for the
blank or reference.
• The double-beam design greatly simplifies this
process by measuring the transmittance of the
sample and solvent simultaneously. The
detection electronics can then manipulate the
measurements to give the absorbance.
05/21/2024 93
Advantage of Single Beam
Instrument
It is well suited for quantitative absorption
measurement at single – wavelength type.
Here, simplicity of the instrumentation, low cost,
and ease of maintenance offer distinct
advantages.
Advantage of Double beam instrument