Download as pptx, pdf, or txt
Download as pptx, pdf, or txt
You are on page 1of 19

Unit V Spectroscopy

Dr. M. A. Sabitha
Assistant Professor in Chemistry
Department of Chemistry
Sadakathullah Appa College
Tirunelveli

1
Electromagnetic Region:
• Radiations such as ultraviolet, visible, infrared, X-rays, radiowaves etc.
• Arrangement of EMR in the order of increasing wavelength is known as electromagnetic
spectrum.

Region Spectra Wavelength nm Frequency Hz


Radiofrequency NMR 100 3x106
1 3x108
ESR
10-1 3x109
Microwave Rotational 10-4 3x 1011
IR Vibrational 7.8x 10-7 3.8x 1014
Electronic Visible 3.8x 10-7 7.5x 1014
UV 10-8 1.5x 1016
Inner shell electronic X-ray 10-11 3x 1018
10-13 3x 1020

Nuclear reaction Γ-ray

2
3
Spectroscopy:
• Study of interaction of electromagnetic radiation with matter.
Molecular Spectroscopy:
• Study of interaction of electromagnetic radiation with molecules.
• Spectrum is called molecular spectrum.
Types of Molecular spectra:
The three types of molecular spectra are:
• Pure rotational spectra
• Vibrational rotational spectra
• Electronic band spectra
Pure Rotational Spectra
• When a molecule absorbs a lower amount of energy that it makes a transition from one
rotational level to another within the same vibrational level.

4
• Rotational spectra are observable in the spectral region of Far Infrared and Microwaves.
• Also, the energies in these spectral regions are exceedingly small. Therefore, we call rotational
spectra the microwave spectra.
Vibrational Rotational Spectra
• When a molecule absorbs sufficient energy that causes the transition of a molecule from one
vibrational level to another within the same electronic level.
• Therefore, in this case, both rotational and vibrational transition takes place. This is how we obtain
vibrational rotational spectra.
• The vibrational spectra are observable in the Near-Infrared Spectral region. We call the vibrational
rotational spectra the Infrared spectra.
Electronic Band Spectra
• When the exciting energy of the radiation is large enough to cause the transition of a
molecule from one electronic level to another electronic level. This transition is
accompanied by both rotational and vibrational level changes.
• Additionally, for each vibrational transition, a set of closely spaced lines appear.
• Because of the corresponding rotational level changes, these closely spaced lines are
known as bands. Hence, we call it the Electronic band spectra. 5
Colorimeter:
• Deals with the measurement of light absorbed by colored substances in solutions.
• The variation of color of a system with change a concentration of some component forms
the basis of colorimetric analysis.
• The intensity of the color may be compared with that of the known amount of the substance.
• The instrument which measures the intensity of the color is known as Colorimeter.
Types of colorimetric analysis:
Two types- Visual colorimetry and Photoelectric colorimetry.
Visual colorimetry:
• Simple visual technique.
• Human eye as detector and brain as amplifier and signal indicator.
• Comparison of samples with standard samples until a match is obtained with Nessler’s tube
as the comparison cell and day light as light source.
• Standard series method, Duplication method, Dilution method and Balancing method.
6
Disadvantages:
• Availability of series of standards.
• Eye will not be able to match the colour in the presence of second coloured substance in
the same solution.
• Eye not sensitive to small differences in absorbance as a photoelectric device.
• Concentration differences less than 5%- difficult to detect.

Photoelectric colorimetry:
• Colorimetric analysis with eye is replaced by photoelectric cell.
• Light of narrow wavelength range obtained by passing white light through filters- source
of radiation –also known as filter photometers.
• Colorimeters using prisms and gratings – spectrophotometers.
• Spectrophotometers are classified- IR, UV, Visible or X-ray spectrophotometer on the basis
of the wavelength region used.
7
Applications of Colorimeter:
• It is used in laboratories and hospitals to estimate biochemical samples such as urine,
cerebrospinal fluid, plasma, serum, etc.
• It is used in the manufacturing of paints.
• It is used in textile and food industry.
• It is used in the quantitative analysis of proteins, glucose, and other biochemical compounds.
• It is used to test water quality.
• It is used to determine the concentration of haemoglobin in the blood.
UV Spectroscopy:
Theory:
• Molecule is exposed to UV or visible light, electron of the molecule is excited from lower level
to excited state.
• Energy required for transition depends on energy difference between ground and excited state.
• Separation between two successive electronic level is very large compared to vibrational level.

8
• Vibrational and rotational transitions take place during electronic transitions.
∆ETotal = ∆Eelec + ∆Evib + ∆Erot

9
σ- σ* transition:
• Transition from bonding σ orbital of molecule to higher energy antibonding sigma orbital.
• Transition corresponds to higher excitation energy, since σ bond is strongest.
• Observed in saturated hydrocarbons and bands appear in vacuum UV region (<200 nm).
Methane-λmax= 122 nm, ethane- λmax= 135 nm.

10
n- σ*:
• Transition from non-bonding ground state to anti-bonding sigma orbital.
Saturates halides, amines, alcohols.
• Methyl alcohol- λmax= 174 nm, methyl chloride- λmax= 169 nm.
π- π*:
• Transition from pi bonding orbital to pi* antibonding orbitals.
• Unsaturated compounds- alkynes, alkenes, carbonyl compounds.
• 1, 3-butadiene- λmax=217 nm, Acetophenone- λmax=240 nm.
n- π*:
• Transition from non-bonding pi orbital to anti-bonding pi* orbital.
• Saturated aliphatic ketones and aldehydes.
• Acetone- λmax=270 nm, Acetaldehyde- λmax=293 nm.
11
Single Beam Spectrophotometer:
• Radiation from source pass through the sample and then through the entrance slit of
monochromator.
• The transmitted radiation is then passed through NaCl prism.
• It is passed through exit slit and then through filter.
• The beam is focused on the detector.
• The detector measures the intensity of radiation.
• According to Beer-Lambert’s law,
• It/ Io = transmittance
• Io- Intensity of radiation is measured when is no sample in the light beam. It- Intensity of
radiation is measured when is no sample in the light beam, knowing Io and It at different
wavelengths, transmittance is determined.
Double Beam Spectrophotometer:
• Source beam is separated into sample and reference beams.
• Reference beam passes through reference cell (solvent) and sample beam passes
12
through sample cell.
• The radiation is passed through exit slit and filter. Beam is focused on detector. The
signal from the detector is amplified and recorded.

13
Double Beam Spectrophotometer

14
Applications of UV spectroscopy:
• Quantitative determination of compounds using Beer-Lambert’s law.
• Measuring intensities of characteristic absorption peaks from time to time, the rate of the
reaction is calculated. Reaction intermediated can be characterized in a few reactions.
• Benzene and iodine in molar ratio of 1:1, new absorption band is formed due to charge
transfer complex between donor benzene and acceptor iodine.
• Distinguish between conjugated and non-conjugated dienes. Acyclic conjugated dienes
show π- π* absorption band more than 217 nm. 1,3- penta diene- λmax=223 nm, 1,4- penta
diene- λmax= 182 nm.
• When number of conjugated double bonds increases, λmax shifts to longer wavelength.
Ethylene- λmax= 165 nm, 1,3- butadiene- λmax=217 nm, 1,3- pentadiene- λmax=223 nm,
1,3,5-hexatriene- λmax= 254 nm.

λmax = 296 nm λmax = 280 nm

15
A1 λmax = 325 nm, A2 λmax = 287, 351 nm

16
IR Spectroscopy:
Principle:
• Energy required to produce vibrational change (0.1 eV) is greater than energy required to
produce rotational changes (0.005 eV).
• So, during vibrational change rotational change also occurs.
• Energy absorbed or emitted is the sum of the two energy changes.
• Spectroscopy is known as vibrational-rotational spectroscopy.
Applications:
Hydrogen bonding:
• Ethyl alcohol in CCl4 gives two absorption bands at 3360 cm-1 (broad) and 3620 cm-1
(sharp).
• Band at higher frequency is due to free OH and other band is due to strong hydrogen
bonded OH.
• Solution is diluted, more intensed free OH band than hydrogen bonded OH.
• Hydrogen bonding weakens the OH group and lowers the frequency of vibrations. 17
• Used to distinguish between intermolecular and intramolecular hydrogen bonding.
• Intermolecular hydrogen bonding decreases with dilution whereas intramolecular hydrogen
bonding has no effect.
Aldehyde and ketone:
• Aldehyde γC=O stretch at 1740-1695 cm-1. Acetaldehyde- 1730 cm-1 and benzaldehyde-
1710 cm-1.
• Ketones γC=O stretch at 1725-1700 cm-1. Acetone- 1725 cm-1, Acetophenone- 1683 cm-1.
• Acetophenone has C=O in conjugation with C=C. Hence absorption is at lower
wavenumber region.

18
19

You might also like