Professional Documents
Culture Documents
LCMS 2020 Introduction
LCMS 2020 Introduction
Introduction
LCMS Group
Shimadzu Europa GmbH
Overview of LCMS
Ionization in LCMS
Vacuumsystem
Ion Optics
Quadrupole Mass Analyzer
Measuring modes
Mass Calibration
Detector
Keypoints for LCMS
Resolution
Signal to Noise
Overview of LCMS
Overview of LCMS
Ionization- Mass-
Detector
source analyzer
Overview of LCMS
Ionization source:
Ions are generated at atmospheric pressure
Mass-analyzer:
Ions are transferred from atmospheric pressure into
the vacuum chamber for mass analysis. Here they are
separated by the mass/charge (m/z) ratio.
Detector:
Ions are detected by Electron-Multiplier
Overview of LCMS
4,5 kV
++++ + Coulomb
+ + + +
+ + ++
+++ + Explosion
Nebulizing gas
N2 (1-1,5 L/min)
1. Charge of ESI-pipe
2. Mist creation of multiple charged droplets
3. Minimizing surface/charge-ratio Coulomb explosion
4. Fully desolvated single and multiple charges ions are created.
Advantages of ESI
Basic compounds:
ionization works better in positive mode
add acidic compound
M-NH2 + AH → [M-NH3]+ + A-
Example: A-=HCOO- AH=HCOOH (Formic acid)
Acidic compounds:
ionization works better in negative mode
add basic compound
Specification-control-sample:
Papavarine:
Measurement mode: Positive
Possible ion-adducts:
Possible solvent-adducts:
Specification-control-sample:
para-Nitrophenol:
Measurement mode: Negative
Possible ion-adducts:
Possible solvent-adducts:
3.0e6
16953.2
2.0e6
998 0.0e6
500e3
+13 16800 16900 17000 Mass
1305
943
400e3 +12
1414
893
300e3 Deconvolution
+20 +11
200e3
1542
849
100e3 +10
1696
1064 1136 1418 +9
0e3
750 1000 1250 1500 1750 m/z
Deconvolution
A : (n) charged
Intensit
B : (n-1) charged
y
A = (M + (n) x 1,0078) / n
B = (M + (n-1) x 1,0078 / (n-1)
M = (A – 1,0078 ) x (n)
m/z n = (B – 1,0078) / (B-A)
MW of Hydrogen: 1,0078 Da
Multiply charged ions
Formula of Myoglobin: C769 H1213 N211 O217
M 16.950,5587; [M+H]+ 16.951,5667
N2 (10-20L/min)
0,0 kV
Heating 500ºC Solvent
charge Charge Transfer
++
++
++
++
Heating 500ºC
4,5 kV
N2 (1-1,5 L/min)
Nebulizing gas
Power supply
APCI: electric field
Positive mode:
Negative mode:
- By using drying gas and heat in the desolvation line these clusters are
destroyed
Reaction-mechanism in APCI
N2 e N 2 2e
N 2 2N2 N 4 N2
N 4 H 2O H 2O 2N 2
Positive mode
H 2O H 2O H 3O OH
2OH H 2O2
H 2O2 H 2O H 2
H 3O M M H H 2O
Negative APCI ionization
O2 e O
2
O n H 2O O H 2O n
Negative mode
2 2
4 HO 3O2 2 H 2O
2
O ( H 2O) n M M H HO
2 2
Factors affecting APCI response
H 3O RNH 2 R3 N R3 NH H 2O RNH 2
Advantage:
DUIS can be a helpful tool especially for unknown
substances and mixtures as it delivers ESI and APCI
information in one experiment
Disadvantage:
DUIS offers slightly reduced sensitivity compared to
single ESI or APCI. (About 30% loss of sensitivity
dependent on application)
APPI
Atmospheric Pressure
Photoionization
APPI-mechanism
Drying and nebulizing gas (N2) To the Analyzer
Solvent (MeOH, ACN, …)
Drying gas
Analyte-molecule
Drying gas N2 (10-20L/min)
N2 (1-1,5 L/min)
Nebulizing gas
N2 (10-20L/min)
Heating 500ºC
N2 (1-1,5 L/min)
Nebulizing gas
Direct APPI
+H
M+H+
M+
UV
(hν=10 or 10.6eV)
Protic solvent
Analyte
Dopant APPI
Molecular Weight +
M+
e-
100,000
M+
+H
M+H+
Dopant M+
UV
Solvent (hν=10 or 10.6eV)
Analyte
Non-Polar
Energetics for Photoionization
Molecular Weight
100,000 Krypton Lamp 10&10.6eV
Ionization
Solvent Ionization
Potentials(IP)
Potentials(IP)
Anthracene 7.4eV Works without
Toluene 7.4eV Dopant-molecules
Fluoranthene 7.8eV
Acetone 9.5eV
Caffeine 8.0eV
4-Nitrotoluene 9.5eV Works only with
Methanol 10.85eV Dopant-molecules
Acetonitrile 12.19eV
Water 12.61eV
Non-Polar
Scope of API-MS
Molecular Weight
10,000
APCI
1,000 APPI
1000
100
Polarity
Comparison ESI / APCI / APPI
Sample stability Good for many Not good for heat- Not good for heat-
analytes unstable analytes unstable analytes
Vacuumsystem
Why do we need vacuum?
1.At atmospheric pressure too many air-molecules (O2, N2 and Ar) are in
the ion-path. That means, the air-molecules disturb the flow-path of the ions.
In case of collision, the ion-pathway is changed and then they collide with
the electric parts (ion-lenses, octapole, quadrupole). After the collision the
ions are lost.
2.The air-molecules are moving in the system and would create a big noise-
ratio, because some of them hit the detector.
Side stage
28 m3/hour 155 L/sec Main stage Dual inlet TMP
155 L/sec
Edwards EXT200/200H
Atmospheric
pressure
Leak valve
Rotary Pump
Edwards E2M28
Dual Inlet TMP
5*10-4 Pa
0,2 Pa
Complete TMP is in use for the
Only half of
analyzer chamber
TMP is in use
for the lens
chamber To roughing pump
Vacuum Pressure Vacuum-Pump-curve
Time
Ion Optics
Ion optics
DL
Without Ion-optics only a few Ions can reach the detector (<<0,1%)!
Ion optics
Why is it necessary to use ion-optics in MS?
Example: With Ion-optics
Skimmer
Q-Array (Focusing)
Octapole (Aligning)
DL
Mass analyser:
Single stage quadrupole
n
Transmissio
Ions can reach No Ions can reach the
the detector detector
m/z-ratio m/z-ratio
Low RF- and High DC-voltage High RF- and Low DC-voltage
n
Transmissio
n
Transmissio
m/z-ratio m/z-ratio
High- and low pass-filter
n
Transmissio
m/z-ratio
When RF- and DC-voltage is set at the quadrupole, so that only one kind of m/z-
ratio can pass the Quadrupole, the ions have only a transmission in a small triangle.
+[U+V cos(ωt)]
+
-++
U V
0
-+- -+ -
-
-
++
+ -[U+V cos(ωt)]
0
U
V
-
Quadrupole and Mathieu-diagram
DC-Voltage
RF unstable
DC stable
RF stable
DC unstable
RF stable
DC stable
RF-Voltage
DC-Voltage
RF unstable
DC stable
RF stable
DC unstable
RF stable
DC stable
RF-Voltage
DC-Voltage
RF unstable
DC stable
RF stable
DC unstable
RF stable
DC stable
RF-Voltage
RF unstable
DC stable
RF stable
DC unstable
RF stable
DC stable
RF-Voltage
A-value Mathieu-diagram
Fixed values:
Used Voltages:
r = 3,9 mm
m/z = 10 U = 2,685 V V = 16 V
f = 1,2 MHz
m/z = 1000 U = 268,5 V V = 1600 V
e = 1,602 *10-19 As
m/z = 2000 U = 537,0 V V = 3200 V
m = Da * 1,66*10-19 Kg
Mathieu-diagram
Calculation of the Peak-tip
A-value
A = 0,240
Q = 0,714
Q-value
Why is the A- and Q-value so important?
For each m/z-ratio the A/Q-ratio is same!
e.g. m/z~2000
m/z~400
m/z~100
m/z~1000
RF-Voltage
DC-Voltage
Why does mass-peaks have a width?
Increasing m/z
RF-Voltage
Intensity
m/z-range
Mathieu-diagram and Resolution
Inten. (x1,000,000)
Int. Max = 2.4 mio
DC-Voltage
609.2
609,2
1,50
Unit Resolution
R~2M
1,25
1,00
0,75
610.2
0,50 610,2
0,25
611.2
611,2
0,00
608,5 609,0 609,5 610,0 610,5 611,0 611,5 612,0 612,5 613,0 613,5 m/z
Inten. (x100,000)
609,3
Int. Max = 1.2 mio
609.3
7,0
6,5
5,0
4,5
610.3
4,0
3,5
3,0
610,3
2,5
611.3
2,0
1,5
1,0
0,5
611,3
612.3
0,0
609,0 609,5 610,0 610,5 611,0 611,5 612,0 612,5 613,0 613,5 m/z
RF-Voltage
7,0
Inten. (x1,000,000)
Int. Max = 11.0 mio
609,1
6,0
609.1
low
Goodworking
Bad workingline:
line: 5,0
4,0
Resolution
Resolution is < ~ 2M
> 3,0
R <2M
Intensity
very goodisspecificity
fine
good 2,0
1,0
No
good
Intensity
specificity
specificity
is worse 0,0
607,0 607,5 608,0 608,5 609,0 609,5 610,0 610,5 611,0 611,5 612,0 612,5 613,0 m/z
Measuring modes
Scan-function
1. Scan
2.
7.
3.
4.
5.
6.
DC-Voltage
Time-range
1 sec
Intensit
Scanspeed
Intensit
y
y
time time
m/
m/
z
z
time time
Intensit
m/
z
y
time
Time
z
m/
Intensit
y
Time
Scan-Mode 15000Da/s
m/
z
m/z=10-2000 Da
Time
ZOOM
m/
time
z
nt
I SD
rem e
I SD
eas u
M
Time
ement
ement
ement
time m/z=100-110 Da
time
IS IS IS
time
Measur
Measur
IS
m/
D D D
Measur
D
z
Mass-range and Scan-speed
Scanspeed without ISD: 15000 Da/s Scanspeed without ISD: 15000 Da/s
m/z: 10 Da – 2000 Da m/z: 100 Da–110 Da
Minimum Event time: 0,1375 s Minimum Event time: 0,0055 s
DC-Voltage
if acquisition time per Da < time of flight in the rods
Novel approch:
1. Changing Prerod-Bias voltage during a
scan to accelerate bigger ions!
2. Changing the GAIN-voltage (slope) on the
Quadrupole (open Quad for bigger Ions)
RF-Voltage
Scan- & Profile-mode
Profile-mode:
- Peak-shape in MS-spectra
- Necessary for deconvolution
- Bigger datafiles (x5 – x10)
- more information than Scan-mode
Scan-mode:
- Also named as centroid-mode
- Used as standard
- Deconvolution not possible
- Smaller datafiles than Profile-mode
Voltage
SIM-Mode (Single Ion Monitoring)
m/z=1600
m/z=609
m/z=82
time
z
m/
m/z=1600
Intensit
m/z=609
y
m/z=82
time
SIM-Mode
Theory (no Inter-Scan-Delay)
Voltage
Time
Praxis (with Inter-Scan-Delay – fix: 5ms)
Voltage
Voltage
Time Time
Inter-Scan-Delay: 5ms Inter-Scan-Delay: 5ms
Measurement-time: 5ms Measurement-time: 1ms
SIM-Mode
Voltage
100 Da/s
500 ms measurement-time
500 ms Inter Scan Delay
50% measurement-time
Inter-Scan-Delay: 5ms Time
Measurement-time: 5ms
Voltage
166 Da/s
Inter-Scan-Delay 5 ms 50 % 5 ms 83,4 %
Measurement-time 5 ms 50 % 1 ms 16,6 %
Time per Da 5 ms 1 ms
As more data points per peak, as worse the statistic per data point
Mass Calibration
Mass calibration of Quadropole MS
m/z
RF-Voltage Voltage
PPG (m/z>1500)
PEG 1000
PEG 600
Raffinose (m/z=503)
PEG 200
Voltage
Detector
Detector
-10 kV
NEGATIVE!
Ion detection – negative Ions
+10 kV
POSITIVE!
Ion detection
Signal-Intensity
Detector voltage and amplification
108
amplification
106
104
Phosphates: are not volatile, will create clogings and ion suppression
Polarity of solvent
ESI-solvents:
A mobile phase containing organic solvent is
preferable for promoting ion vaporization and formation of
fine charged droplets in atmospheric
pressure.
Under normal reverse phase column conditions,
acetonitrile/water achieves higher ionization efficiency
than methanol/water. The pH also has a great effect.
Mobile phases and LC/MS
• Fragmentation (Q-Array-Voltage)
• DL-temperature
• APCI-temperature
• DL-voltage
• avoid frequent auto tuning.
calibrate every 3 - 6 month.
• Methanol, Isopropanol, water and
! Formic acid ! for cleaning
Troubleshooting
R = Resolution
m = mass of the 1. peak
Δm = detectable mass-difference
FWHM = 0,7 R=m/z / FWHM
50% Peakheight
10% Peakheight
Resolution Quadrupole
R = 2M
two peaks with 1Da difference and an
FWHM = 0,7 overlap at 10% Peakheight are
separated.
Same FWHM over complete mass-
range
Resolution increases with m/z ratio
Resolution IT-TOF
FWHM = 0.1 FWHM = 0.2 FWHM = 0.4 FWHM = 0.5 FWHM = 0.7
Quadrupole-range
+3 RMS
Average
-3 RMS
min
max min
NOISE
2
RMS = Root-Mean-Square Deviation = σ
That´s it
Questions?