Professional Documents
Culture Documents
Scintillation Camera Testing
Scintillation Camera Testing
Testing
• Any PET
radionuclides?
• Any
radiotherapy?
Gamma Camera
Installation Tests
• Head Leakage
• Collimator Hole Alignment
• Couch Alignment (SPECT)
• Linearity
Head Leakage
• To find defects in head shielding
• Move radioactive source around on
external surfaces of camera head
during acquisition
• Leaks at support points
• Visual test
Collimator Hole Angulation
• Point source (150 MBq) as far away
from collimator as possible
Frequently limited by room design
• Not easy to radiograph collimator
• Look for areas where holes are not
properly angulated
• Visual test
Couch Alignment
• SPECT is very dependent on how far
the head is away from the patient.
Errors in table alignment degrade image
quality and could cause patient
collisions
• Should be aligned to within 1 cm
Linearity
• Visual with bar phantom
Part of QC but not good enough for
acceptance testing
• Special NEMA phantoms
Request from manufactures as part of
purchase
• NEMA verification phantom
Linearity Verification
• Software determines
center of each hole
and checks linearity
Yc = N iYi/ N i
6
2
2
0 2 4 6 8
1
Intrinsic Uniformity
• Image of point
source without
collimator
• 5 Diameters ~1%
• ~2.5 m h
D
2.5 m
5 5
4544 .9 553
.2 90 2.7
36 73
.8 74
.6
07
68
4 4
181
11335.0
165
2038
11058.3
2
5.8
8
2
3.8
1
1849.1
6 0.
9.4
8.
0
2
59 90
2.
2.
2280.6
12
74
13
86
14
15
3 3
.4 .4
71 16
90 92
2 2
1 1
1 2 3 4 5 6 1 2 3 4 5 6
1 1
• FWHM of
photopeak
• Spectrum may
not be easily
available
• FWHM depends
strongly on FWHM = 15/122 =12%
scatter
10% is typical specificaton
Energy Errors
• Spatial Location vs gamma
energy
Ga-67 for example
• 0.5 mm typical
E1 E2
Multiple Window
• Ga-67
91 keV
185 keV
300 keV
• Image
Compare Centroids
1.
7
5 5 1.
8
1.
9
2.0
4 4
2
2
3 3
2.1
1.
7
2 2
1.
6
1 1
1 2 3 4 5 6 1 2 3 4 5 6
1 1
R1 R2
p =[2R12/(R1+R2)2]/ln[(R1+R2)/R12
Background
Source 1
Source 12
Source 2
Source 2
Source 12
Source 1
Deadtime
• Typical Results
4.1 usec clinical
window
1.9 usec wide window
• Deadtime can be
decreased with
sacrifice of image
quality
Image Quality
• Image quality for clinical examination
depends upon
Capabilities of camera
Parameters used for imaging
Patient body habitus and physiology
• There are a wide variety of “image
quality” phantoms
Intrinsic Intrinsic
Linearity Energy Cor
Energy
Resolution
Intrinsic Intrinsic
Deadtime
Uniformity Resolution
Imaging
Technique Image
Quality
Image Quality Phantom
• Big Picture
• If poor image
quality we still
have to find the
cause
loss of resolution
poor techinque
etc
General Safety Checks
• Electrical Issues
• Locks
• Collision Sensors
• General Condition of
Room
• General Radiation
Safety Checks
SPECT Performance
Phantom
• Care must be taken to
reproduce the set up precisely
Activity
Placement of phantom
Distance of collimator from
phantom
Technique
Reconstruction
Routine Quality Assurance
• System Flood – Daily
• Resolution & Linearity – Weekly
• High Count Floods (As Required)
• Center of Rotation – Monthly(?)
• Monitor Performance (Monthly)
• Printer Performance (Monthly)
• SPECT Performance - Quarterly
Daily System Flood
• Visual Assessment
• Quantitative Assessment
Reduced Performance
14%
12%
10%
8%
6%
4%
2%
0%
Jan-93 Oct-95 Jul-98 Apr-01 Jan-04
UFOV CFOV
Resolution & Linearity
• Visual Assessment
• If using 4 quadrant
bar rotate weekly
Center of Rotation
• As necessary -
usually monthly
Center of Rotation
• SPECT assumes heads always look at
some constant rotation point
COR
• If heads are mis-aligned or have some
mechanical motion
• Image will have blurring and circular
artifacts
• COR must be tested for and corrected
for
• SPECT of 3 point sources
Monitor Performance
• Brightness (Photometer)
• Contrast
• Resolution
SMPTE Test Pattern