Professional Documents
Culture Documents
RADIATION DETECTORS-parikshit
RADIATION DETECTORS-parikshit
As charged particle passes through the chamber the interact with the
medium (supersaturated vapours alcohol or water , superheated lq H), and
create ionisation tracks which are then recorded
Most of the modern day detectors like gas filled detectors and
semiconductor detectors work on the principle of recording ionisation
created by radiation in the detectors in the form of an electrical signal.
WILSON CLOUD CHAMBER BUBBLE CHAMBER
GENERAL PROPERTIES OF
DETECTORS
SIMPLIFIED MODEL OF
DETECTOR
Charge
created
Ejection of within the
high speed detector is
electrons , and collected
they deposit producing an
their energy electrical
Interaction of signal
radiation by producing
within the charge within
active volume a volume
of the detector
SENSITIVITY
Capability to produce a signal corresponding to an event
Factors affecting the sensitivity to a given type and E of rad:-
CURRENT MODE
BLOCK DIAGRAM :-
It is extension of current mode in which we measure
only the fluctuation component σ(t) by introducing a
circuit which blocks the average current and allows σ(t)
to pass.
This mode enhances the relative response of the
detector.
Useful in when making measurements in mixed
radiation .
Used in nuclear labourites.
PULSE HEIGHT
SPECTRA
Pulse amplitude distribution is a fundamental property
of the detector output which is used to deduce info
about the incident radiation
In spectroscopy systems the objective is to sort each
pulse according to its amplitude, the pulses recorded
are sorted into the channels matching their amp and
hence PHA is obtained
A) Differential pulse height spectra
B) Integral pulse height spectra
Both of the curves convey the same info ,amp of any
pulse height is given as the absolute slope in the lower
curve.
RESPONSE FUNCTION
Ideally it should be a Dirac Delta function ,ie for a fixed
energy the output has a single fixed value
Linear response means , pulse height spectra corresponds
directly to energy spectra
Determination
- Interaction type, material and geometry
a) Is response function for 661 kev , for Ge detector large
photoelectric cross section hence larger photo peak,
relatively small continuous Compton spectra.
b) Organic scintillator , low z hence more Compton
scattering
ENERGY
RESOLUTION
Is a measure of detectors response function
It is defined in terms of
Larger width represent large amt of fluctuation
Causes of fluctuations
-Drift form op characteristics
-Random noise from inst
-Statistical noiseFANNO FACTORnature
due to discrete
the measured signal
DEAD TIME
The minimum time between two pulses so that they are
recorded as 2 different pulses .
• PARALYZABLE
• NON PARALYZABLE
Non paralyzable Paralyzable
Also known as non extendable dead Also known as extendable dead time.
time. Arrival of a second pulse during the
The detector remain insensitive dead period extends this period by
while processing an event . adding its dead time starting from
The detectors dead time remains the moment of its arrival.
fixed and is not hampered by the Model :-
events occurring during this period
Model :-
− 𝑛г
𝑛=
𝑚
1 −𝑚 г
𝑚=𝑛 𝑒
TYPES OF
DETECTORS
Ionisation Excitation
detectors detectors
Chemical
Other detector
detectors
GAS FILLED DETECTORS
RESPONSE CURVE
Pulse height
Voltage
1. IONIZATION CHAMBER
Excitation:- X + p +p
Ionization mechanism
1. Primary ionization
2. Secondary ionization
3. Penning ionization
Ne* + Ar Ne + Ar+ + e
4. Formation of molecular ions
He+ + He He2+
W-Value
No. of ion pairs formed is determined the W-value which is average energy
expended to create an ion pair
Loss of charges
ELECTRON ATTACHMENT RECOMBINATION
Arises from the fact that ion pairs are An encounter b/w an e/ion after
formed along the track of ionizing they have left the site of formation
particle
Increases with increase in dose rate
Increases with increase in LET
Charge separation and collection
Independent on the rate at which the should be rapid to minimize
tracks are formed . recombination
Ionization current
Measurement of ionization current is the basic
principle of ion chamber
FACTORS AFFECTING SATURATION CURRENT
1. INCREASE IN HUMIDITY
2. RECOMBINATION -columnar and volume
3. PERTUBATION in measured current due to a
difference b/w the direction of diffusion and
drift current
-ΔI/I
The size of this loss in the saturation current
depends on applied voltage and the size of the
quantity E
DESIGN AND CONSTRUCTION
1. INSULATORS AND GUARD RINGS
Supporting insulators are provided b/w the 2
electrodes.
They are made up of a high resistive material to
minimize leakage current
Leakage current due to absorbed moisture or other
contaminants is always present on the surface of the
insulator
Most of voltage drop occur across the outer
insulator ,leakage current in this case do not pass
through the measuring instrument
Voltage drop across the inner segment will be equal to
that across the ammeter.
2. Measurement of current
ELECTROMETER ELECTRET
Measures the current by sensing the Group of solids that have a permanent
voltage drop across a series resistance ,this electrical polarization are called electret
voltage signal is then amplified
Potential difference b/w the surface of
Weakness – requires dc coupling electret
throughout
Placed with one surface in contact with
DYNAMIC CAPACITOR/VIBERATING REED the conducting wall, creating an E in the
cavity
Converts the dc signal into ac which is
more suitable amplification (by varying the Ion pairs collected serve to neutralize
capacitance about the avg value) the charge carried by the electret and
the potential drop is measured in the
electret
Application
Survey meters
Ionization chambers are widely used in the radiation industries as they provide
an output that is proportional to radiation dose.
They find wide use in situations where a constant high dose rate is being
measured as they have a greater operating lifetime than standard Geiger-Müller
tubes, which suffer from gas break down and are generally limited to a life of
about 1011 count events.
Additionally, the Geiger-Müller tube cannot operate above about 104counts per
second, due to dead time effects, whereas there is no similar limitation on the ion
chamber.
Medical radiation measurement
In medical physics and radiotherapy, ionization chambers are used to
ensure that the dose delivered from a therapy unit
or radiopharmaceutical is what is intended.
Chamber types
Thimble chamber
Most commonly used for radiation therapy measurements is a cylindrical
or "thimble" chamber.
The active volume is housed within a thimble shaped cavity with an
inner conductive surface (cathode) and a central anode. A bias voltage
applied across the cavity collects ions and produces a current which can
be measured with an electrometer.
Parallel-plate chambers
Parallel-plate chambers are shaped like a small disc, with circular collecting
electrodes separated by a small gap, typically 2mm or less.
The upper disc is extremely thin, allowing for much more accurate near-surface
dose measurements than are possible with a cylindrical chamber.
Monitor chambers
Monitor chambers are typically parallel plate ion chambers which are placed in
radiation beams to continuously measure the beam's intensity.
For example, within the head of linear accelerators used for radiotherapy, multi-
cavity ionization chambers can measure the intensity of the radiation beam in
several different regions, providing beam symmetry and flatness information.
Smoke detectors
The ionization chamber has found wide and
beneficial use in smoke detectors. In a
smoke detector, ambient air is allowed to
freely enter the ionization chamber.
The chamber contains a small amount of
amrecium-241, which is an emitter of alpha
particles which produce a constant ion
current.
If smoke enters the detector, it disrupts this
current because ions strike smoke particles
and are neutralized. This drop in current
triggers the alarm
ADVANTAGES DISADVANTAGES
Can be made into different sizes and Less sensitive as the charge or current
shapes formed is very small.
Since no of ion pairs formed is a fn Owing to the small output the effect
of energy deposited ,response is of external condition can effect the op
directly prop to the dose rate to a measureable,.i.e anything that can
change the density can effect the
Is used to measure the X as it is
response particularly in thin wall
defined in air ,when used with photon ion ,vented chambers op dec by 2%
rad gives the actual exposure rate with temp inc of 10 deg, and dec by
2.3% for every inch of hg decrease
2.PROPORTIONAL COUNTERS
A proportional counter utilizes the
phenomenon of gas
multiplication to increase the pulse
size
Multiplication is a consequence of
the motion of a free electron in a
strong electric field .i.e when the
applied E>Eth
Multiplication model :-
Working
Types of chamber
Cylindrical (constructed using self contained gas )
Gas flow(constructed using continuously cleaning gas )
Design features of
cylindrical chamber
Choice of
Sealed tubes Fill gas
geometry
1. Choice of geometry
1.Cylindrical geometry
Gas multiplication requires large values of E ,which is attainable in cylindrical
geometry
Region of gas multiplication should be small so that most of the e are formed outside
this region , hence each e undergoes same multiplication regardless of its original
position
2. Sealed tubes
Factors affecting proportionality
A. Axial distortion of E, due to non uniform thickness of anode wire
B. Distortion of E near the ends ,and is corrected by using FIELD TUBES
Fill gas
Gas is chosen do not exhibit an appreciable electron attachment coefficient (as the
multiplication critically depend on the migration of e)
Gas multiplication is based on secondary e collision with neutral atoms in active
region
Small amount of additional gases are added to reduce
A. Penning effect B. Quenching gas
Type of radiation To measure the low energy X rays It can detect the
that can be gamma ray alpha particle , alpha , beta
detected exposure , survey beta particle, particles , low
meter neutrons can be energy and high
detect and energy X rays but
differentiate d can’t differentiate
Use in In pulse mode can low energy X rays Can’t be used
spectroscopy be used for
spectroscopy
Dead time 50-100 micro sec.
- -
http://www.scionixusa.com/pages/navbar/scin_crystals.html
ORGANIC SCINTILLATOR
FEATURES
MECHANISM
EFFICENCY =fraction of all incident particle
energy converted into light
QUENCHING=rad less de excitation
In all most all organic scintillators excitation
energy undergoes substantial transfer from
molecule to molecule before de excitation
TYPES OF ORGANIC SCINTILLAOR FEATURES,EXAPMLES
FEATURES
Conduction band , valence band ,
activator site (impurities).
MECHANISM
e hole pair production
Moving of e to conduction band
Recombining of e at activator sight
resulting in emission of photon in the
visible region.
TYPES FEATURES
1.NaI(Tl) Excellent light yield, standard for routine g spectroscopy , poor res at higher temp
2.CsI(Tl) and CsI(Na) Large g abs coefficient/size comp to NaI, used in space application , less brittle, less
hygroscopic ,
Geophysical, general radiation detection
1. BGO Major ad large density (7.13g/cm3) and z, rugged in nature, low light yield, no long
decay compo hence no after glow,
Costly,
Particle physics, geophysical research, PET, anti-Compton spectrometers
2.CdWO4 Yield is 40% of NaI(tl) ,high p and z, used for g ray spectroscopy, low decay time , Very
high density, low afterglow, radiation hard, DC measurement of X-rays (high intensity),
readout with photodiodes, Computerized Tomography (CT)
BLOCK DIAGRAM
Photomultiplier Tube
Predicted response functions
A. SMALL SIZE
FEATURES
Full energy peak
Compton continuum
Double escape peak
B. VERY LARGE SIZE
FEATURES
Full energy peak
C. INTEMIDEATE SIZE
FEATURES
Full peak
Double /Single escape peak
Multiple comp scatt.
Compton valley
Neutron Detectors
They are detected through nuclear reaction that results in prompt energetic
charged particles such as alpha, protons etc.
Combination of a target to carry out conversion + a conventional detector
TYPES OF NEUTRON
1. Thermal ( 0.4 eV)
2. Intermediate ( 0.4 – 200 keV)
3. Fast Neutrons (200 keV- 20MeV)
4. Relativistic (> 20MeV)
SLOW NEUTRON DETECTION
BORON ACTIVATION FISSION CHAMBERS