Professional Documents
Culture Documents
Parul-beam Direction and Modification Devices Used in Tele
Parul-beam Direction and Modification Devices Used in Tele
Transmission penumbra
Geometric penumbra
Physical penumbra
Transmission penumbra
It is caused due to variable transmission
of beam through non divergent
collimator edge.
Its extent increases with larger
collimator openings due to greater
obliquity of the rays at the edge of jaws.
1 cm or wide strips of copper-tungsten
alloy are placed on inner surface of
jaws.
Density of this alloy is greater then
lead , thus more attenuation is
provided.
Geometric penumbra
It is due to the finite dimensions of source.
• The geometric width of the penumbra () at any
depth (d) from the surface of a patient can be
determined by considering similar triangles ABC
and DEC.
DE=
• Therefore, if AB = s, the source diameter, OM =
SDD, the source to diaphragm distance, OF = SSD,
the source to surface distance, then the penumbra
(DE) at depth d is given by
Physical penumbra
• Another method is to
use secondary blocks
placed close to the
patient ( 15 – 20 cms).
SC1D < SC2D P1
P1 > P2 P2
Multi leaf collimators (tertiary collimators) :
A multileaf collimator (MLC) is a device made up of individual
"leaves" of a high atomic numbered material that can move
independently in and out of the path of a particle beam in
order to block it.
MLCs are used on linear accelerators to provide conformal
shaping of radiotherapy treatment beams for improved
treatment delivery efficiency.
Applications :
1. Less time consuming than conventional methods which
use lead blocks for shaping.
2. In arc therapy, leaves move continuously according to the
shape of tumor at particular gantry angle.
3. Achievement of beam intensity modulation using the
motion of the MLC leaves during irradiation to create a
dynamic compensating filter.
Configuration of MLCs
1st level configurations:
• Replacement of upper jaws
General Electric , has the same max. extension of the leaf across the
centerline as Siemens i.e. 10cm.
The leaves in the Varian collimator travel on a carriage to extend their
movement across the field. Distance between the most extended leaf
and the most retracted leaf on the same side can only be 14.5 cm.
• Extension of leaf across the central line in case of symmetric and
large field size is not possible . But in case of asymmetric fields ,
the Varian carriage can be moved to the field center and a leaf can
be extended 14.5 cm beyond the field center.
Attenuation
Material properties
Pure tungsten with density of 19.3g/cm3 is very brittle
• Isocentric mounting
• Electron applicators
Front and back pointer
Principle :
Any straight line can be defined by two points and this method requires
marking of two points on the patient surface such that the line passing
through these points will pass through the tumor center .
WORKING:
Using radiography the two points are decided such that the line joining them will
pass through the center of tumor .
Movable distance indicator is used as front pointer and designed such that its
adjustable rod lies on the central axis of the beam.
Another rod firmly attached to the head of the machine can act as a back pointer
that points towards the exit point of the beam on emerging central axis from the
patient’s body.
This rod can be moved forward or backward accordingly .
The marks are made on the patient “shell ” made of cellulose acetate sheet or
plastic which can fit accurately and easy to be put on and taken off .
This provides more accuracy and reproducibility .
Pin and Arc
It is another widely used aid for accurate beam direction ,mostly used where exit points may be
inaccessible to the back pointer.
Construction :
• Ruler like scale bar R connected to the x ray tube.
• Sliding frame S and arc T is attached to this frame.
• Centre of arc T of the circle lies on the central axis
at O.
• Carriage U which carries a rod/pin V which can be
clamped at any position .
• If length of V is at lowest position then its lower
end is on the central axis at O.
• d is read off by scaled rod V .
• D is measured by R .
• Angle between V and central axis is measured
using angular scale on T .
WORKING
Depth of tumor d and its center O is localized
below surface -mark M .
The pin V is withdrawn the required distance d
and its lower end is kept at M vertically above
O.
Keeping pin V at that position rest of the
equipment is rotated about O.
Central ray of the beam will always pass
through the tumor center.
At any desired angle distance D is adjusted
accordingly and can be measured using R.
Thus D is the depth of the tumor center from
the surface at any particular field which is
useful for dose computation .
Isocentric mounting
X ray tube or tele -curie source is made
to move in arc around patient about an
axis at 100 cm from focal spot.
treatment table is designed to rise and
fall and also to rotate around axis
vertical to the rotational axis of the
source which cut it at iso-center point.
PRINCIPLE :
o If the tumor is placed at isocentric point
using any method central ray will always
pass through it .
PROCEDURE
Firstly front pointer is drawn back a distance equal to the depth of the tumor
center .
Patient on table is raised equal to that distance so that the tip of the pointer
touches the skin mark .
In this way tumor center is placed at isocenter .
In both cases the rate at the center is greater than the rate out
towards the beam edge
Flattening Filter
• ‘Beam-flattening' filter which reduces the central exposure rate relative to that near the edge of
the beam, to give, as nearly as possible, a constant rate across the beam .
• To achieve this result the filter must be thickest at the center and taper off to nothing towards the
edge
• Copper or aluminium is usually used in making these filters.
• It does not alter the central axis percentage depth dose values. Accurate positioning of the filter is
more important than the choice of material.
Flattening in KV energy beams
• They produced 'flat' isodose curves over a range of depths of about 5 or 6 cm, but not
over the wide depth range .
• To equalize the centre and edge rates , only primary radiations are altered.
6 MV 6.5 cm
Thus, 2n = 1/0.05 = 20
10 MV 7.0 cm
OR, n log 2 = log 20.
25 MV 7.0 cm
n = 4.32
The relationship holds true, only for mono energetic x-ray beams .
Compensators
• Where the beam enters the patient obliquely or
through a curved surface compensation is used.
• More radiation will reach P and Q respectively than
would be indicated by any isodose chart since to
reach those points the beam has suffered less
attenuation
• Primary radiation reaching P and Q will be the same
provided that A’C’ equals AC and D’E’ equals DE
respectively.
• Advantageous to place the compensator at
position II because it does not affect the skin
sparing effect in high energy beams .
Designing of compensators
Three separate stages
1. Replacement of the missing tissue area with a series of
square-section sticks, each having an end section of
15mm2
2. This series is displaced at some distance from the patient
surface towards source. Thus the required end section
area of each stick would be
x= mm
3. Replacement of the unit density material of sticks with
another dense material like aluminium or brass . The
length of the sticks now required will be less to provide
the same attenuation.
• The required thickness of a tissue-equivalent compensator along a ray divided by the
missing tissue thickness along the same ray may be called the density ratio or thickness
ratio (τ = h′/h)
• If, for given irradiation conditions, τ is chosen for a certain compensation depth, the
compensator overcompensates at shallower depths and undercompensates at greater
depths.
• An average value of 0.7 for τ may be used for all irradiation conditions provided d greater
than or equal to 20 cm
• The thickness ratio is used to calculate compensator thickness (tc ) at a given point in the
field
tc = TD · (τ /ρc)
where , TD is the tissue deficit at the point considered
ρc is the density of the compensator material.
Compensators in TBI
• The compensators are usually designed in three pieces: one for the lower
extremities, one for the head and neck region, and one for the lungs.
• The first step in designing tissue compensators is to determine the tissue deficit
(TD),
TD=Lref −L+(1−ρlung )Llung
where Lref is the lateral separation at the umbilicus
L is the lateral separation at that particular anatomical location,
Llung is the separation of the lung determined from the anterior
radiograph
ρlung is the density of the lung 0.25 g/cm3
• The compensator thickness, Lc, is determined using the following equation :
Lc =
θ = 90 – φ/2
To produce a high dose zone which is clinically
acceptable, it is not necessary for the isodose
curves to be strictly parallel.
Physical Non-physical
wedges wedges
Individualized Universal
wedge wedge
Physical wedge
• A wedge-shaped absorber that causes a progressive decrease in the intensity across the beam,
resulting in a tilt of the isodose curves from their normal positions.
• The sloping surface is made either straight or sigmoid in shape; the latter design is used to
produce straighter isodose curves.
• Made of a dense material, such as lead or steel
Manually handled wedges Motorized wedge
• Are mounted on a transparent plastic tray or a • Is placed in the field internally i.e., an
frame that can be inserted in the designated internal motor slides the wedge into
slot in the head of the machine . position.
• A set of wedges is usually employed on each • Consist of a single large wedge i.e. 600
megavoltage machine covering the angles 15, • Placed above the secondary collimating jaws
30, 45 and 60 °. .
• Placed at least 50 cm from the isocenter
• So that it does not destroy the skin-sparing
effect of the megavoltage photon beam
Individualized wedge Universal wedge
• Requires a separate wedge for each beam • A single wedge serves for all beam widths.
width, optimally designed to minimize the •
loss of beam output. Fixed centrally in the beam, while the field
can be opened to any size.
• Thin end of the wedge is aligned with the • Useful for linear accelerator beams where
border of the light field
the output is not fixed and can be varied.
• Preferred for use in cobalt teletherapy due
to constant output.
Dosimetric effects
• Decrease in the dose output due to attenuation is characterized by the wedge
output factor or simply wedge factor, defined as the ratio of doses without and
with the wedge, at a point in phantom along the central axis of the beam.
[setup conditions : (10 X 10)cm field size , measurement depth is 10cm , dose
2
Advantages :
• Automation of treatment delivery
Disadvantages :
• Greater dosimetric complexity in the acquisition of commissioning data
• Effect of Lucite shadow tray on dose buildup for 10-MV x-rays. Percent depth–dose distribution is
plotted for various tray to surface distances (d). 10-MV x-rays, tray thickness = 1.5 g/cm2, field size
=15 × 15 cm2, source to surface distance = 100 cm, and source to diaphragm distance = 50 cm
Modification in electron beam
Scattering foil
• Beam modifying device used in electron beam
therapy
NEED:
• To spread the thin beam of electrons(3mm in
diameter) to cover the treatment volume
• To get uniform electron fluence over the
treatment volume
• Mostly used material for making scattering foil
is lead
• Thickness should be such that mostly electrons
are scattered
Lead cutouts
•
Thickness of lead required :
Lead cutouts : lead cutouts are used to restrict the
thickness
electron fields of leaddesired
to the required
areato stopisthe
which primary
to be
electrons has been investigated by Giarratano et
treated.
al. (1975),
• Lead who concluded that a lead thickness
cut out can be manufactured from commercially
in
millimeters
available sheetsequal
of leadto the most probable electron
energy at the surface (in MeV) divided by two is
• Placed directly on the patient’s surface to define the
treatment areaadequate to provide shielding
An extra millimeter
• Produces a field with sharp edges
of lead can be added to provide an additional
• Should be considered for small fields, low electron
margin of safety.
energies and when
e.g. For critical
20Mev, areas lie directly
thickness adjacent
of Pb = to
desired field.
= 11 mm
Isodose distribution at the edge of lead cutout