Professional Documents
Culture Documents
Yogendra [Defib] 16
Yogendra [Defib] 16
YOGENDRA REDDY
•Pin number:- 19055-BM-016
•Topic:- Defibrillator
•College:- SGPR Govt polytechnic College
DEFIBRILLATO
R
OBJECTIVE
1. Define defibrillator
2. Describe need And history of defibrillation
3. Describe the principle and mechanism Of defibrillation
4. Types and classes of defibrillation
5. Describe the automated external defibrillator
6. Identify the precautions and risks
7. Troubleshooting of defibrillator
DEFIBRILLATOR
INTRODUCTION
• Definition:-
Defibrillation is a process In which an electronic device sends An electric
shock to the heart to stop an extremely rapid, irregular heartbeat and restore
the normal heart beat Rhythm.
Defibrillation is a common treatment for life threatening cardiac
arrhythmia, ventricular fibrillation and pulse less ventricular Tachycardia.
NEED FOR A DEFIBRILLATOR
• High voltages are required for external defibrillation than for internal
defibrillation.
• A corrective shock of 750-800 volts is applied within a tenth of a second.
• That is the same voltages as 500-533no of AA batteries.
NORMAL CARDIAC CONDUCTION
CARDIAC ARREST
• Types of electrodes:-
1. spoon shaped electrode
Applied directly to the heart.
2. paddle type electrode
Applied against the chest wall.
3. Pad type electrode
Applied directly on chest wall.
DEFIBRILLATOR ELECTRODES
ELECTRODE PLACEMENT OF AED
CLASSES OF DISCHARGE WAVEFORM
CLASSES OF DISCHARGE WAVEFORM
• For each minute elapsing between oneset of ventricular Fibrillation and first
defibrillation, survival decreases by100%.
• Defibrillators should be portable, battery operated, small size
• Energy in defibrillators usually stored in large capacitors.
• Total energy stored in capacitor:
Wc=1/2cv2c vc=capacitor voltage
EXTERNAL DEFIBRILLATOR
AUTOMATIC EXTERNAL DEFIBRILLATOR
• Attach the external and internal paddles if the monitor reads,‘‘No paddles’’.
• Check to ensure that the leads are securely attached if the monitor reads,‘No leads’.
• Connect the unit to AC power if the message reads, ‘‘Low battery’’.
• Verify that the energy select control setting are correct if the defibrillator does not charge.
• Change the electrodes and make sure that the electrodes adapter cable is properly
connected if you receive a message of‘‘PACER FAILURE’’ Restart the pacer.
• Close the recorder door and the paper roll if the monitor message reads,‘check recorder’.
THANK
YOU