Professional Documents
Culture Documents
System M Insevice Overview 11.8.11
System M Insevice Overview 11.8.11
Cases. The system continuously measures: SvO2 Hematocrit/Hemoglobin Venous flow Venous differential PO2 Temperature SaO2 Arterial flow Cardiac Index Micro-emboli PCO2
Key Features:
Non-invasive. Can be used on any tubing. Real-time, not a trending device. Accurate with no calibration needed. Portable with one minute set-up and quick release mounting system. No disposable costs.
No Disposables:
Quick Set-up With Easy to Clip On Non-Invasive Sensors Flexibility in Placement of the Sensors On the Tubing No Flow Disturbances Decreased Risk of Infections and Clotting Minimal to No Calibration Needed
Monitor Overview
Flow Connector
Flow Connector
M4 System Sensors:
Venous O2 Saturation Sensor
Flow Sensor and Bubble Detector
CO2 probe
The Parameters Measured By the System M are Considered Critical Care Parameters. Therefore it is critical that changes in these parameters be identified and acted on quickly. The ability to measure these parameters in realtime is therefore an obvious benefit.
Saturation/Hct sensor cable should be connected to the Left side of the monitor. The sensor cable will split into a venous sensor and an arterial sensor.
Attach the required size of pipe-clip on the tubing. Ensure the retention clips are fully engaged.
Insert the sensor between the retention hooks until the sensor is secure.
To remove the sensor, hold the tubing as shown and squeeze the retention hooks until the sensor can be pulled free or the pipe-clip can be removed for re-use.
Ensure that both the tubing and sensor assembly are free of writing, dust, and any form of contamination.
Open the lid and slide the tubing into the sensor.
Warning: Only use Isopropyl Alcohol solution to clean the HCT/S02 sensor.
Do not use water based or petroleum based lubricants. Sensor calibration is determined by tubing dimensions. Using on tubing other than the specified size generally yields inaccurate measurements and may cause erratic zero baseline drift.
As with the saturation sensor the size of the flow sensor must match the tubing size on which it is placed. When flow is at zero the system can be zeroed out for more accuracy.
The Gas Measurement Module is positioned between the Gas blender and the Oxygenator. This sensor monitors the gas delivered to the oxygenator. Ensure an anti-bacterial filter is positioned between the Gas Measurement Module and the oxygenator.
Ensure the gas module is mounted in a horizontal attitude. Mount the gas measurement module via its pole clamp in the orientation shown below. Connect the gas measurement module electrically to the M3 monitor by inserting the lead from the M3 monitor into the connector indicated.
Monitor Connector
Connect the exit CO2 sensor to the Gas out-flow site of the oxygenator. Keep the sensor as upright as possible (as shown in picture) to stop water condensation from entering the sensor.
Connect the exit CO2 sensor electrically to the gas measurement module by inserting the connector into the socket indicated. Do not block or restrict the exit port.
Displayed Parameters
Graphed Parameters
SD Card Information
Stop/Record Button
Graph Controls Parameter and Alarm Set-up Button Patient Details Set-up Button
The SD card does not need to be in the system when doing a case but it can be used to record the case and then downloaded to a spreadsheet at a later date.
When Connecting the Sensors and the Power Cable to the Monitor Always Line the Dot on the Connector to the Dot on the Monitor.
times be handled with care. Do not exceed the minimum bend radius of 50 mm.
Battery power lasts approximately 30 minutes.
Keep the monitor plugged in when not in use in order to prevent draining the battery.
Thank you!